FDA Adverse Event Malfunction Summary report: N

LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (LARGE)

MDR report key: 1123806 · Received August 14, 2008

Report

Report Number
2024601-2008-00429
Event Type
Malfunction
Date Received
August 14, 2008
Date of Event
July 15, 2008
Report Date
July 18, 2008
Manufacturer
COSTA RICA
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

TAPER II. VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS OF THE DEVICE NOTED DAMAGE FROM A SHARP INSTRUMENT TO THE BAND TUBING (THIS IS THE PORTION OF TUBING LOCATED BETWEEN THE STAINLESS STEEL CONNECTOR AND THE BAND, NOT THE STAINLESS TEEL CONNECTOR AND THE PORT). THERE WAS NO INDICATION OF WEAR RELATED DAMAGE TO THE PORTION OF THE LAP-BAND SYSTEM RETURNED. THE LAB WAS UNABLE TO DUPLICATE OR CONFIRM THE REPORTED EVENT. THERE IS NO OTHER INFORMATION AVAILABLE AT THIS TIME FROM THE SURGEON TO DETERMINE THE CAUSE OF THE ALLEGED EVENT. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."

Description of Event or Problem · 1

REPORTED AS A PORT LEAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (LARGE) DEVICE FOR TREATMENT OF MORBID OBESITY LTI COSTA RICA NA 1388131

Patients

Seq Age Sex Outcome Treatment
1 47 YR