LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (LARGE)
Report
- Report Number
- 2024601-2008-00429
- Event Type
- Malfunction
- Date Received
- August 14, 2008
- Date of Event
- July 15, 2008
- Report Date
- July 18, 2008
- Manufacturer
- COSTA RICA
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- PHYSICIAN
Narratives
TAPER II. VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS OF THE DEVICE NOTED DAMAGE FROM A SHARP INSTRUMENT TO THE BAND TUBING (THIS IS THE PORTION OF TUBING LOCATED BETWEEN THE STAINLESS STEEL CONNECTOR AND THE BAND, NOT THE STAINLESS TEEL CONNECTOR AND THE PORT). THERE WAS NO INDICATION OF WEAR RELATED DAMAGE TO THE PORTION OF THE LAP-BAND SYSTEM RETURNED. THE LAB WAS UNABLE TO DUPLICATE OR CONFIRM THE REPORTED EVENT. THERE IS NO OTHER INFORMATION AVAILABLE AT THIS TIME FROM THE SURGEON TO DETERMINE THE CAUSE OF THE ALLEGED EVENT. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."
REPORTED AS A PORT LEAK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (LARGE) | DEVICE FOR TREATMENT OF MORBID OBESITY | LTI | COSTA RICA | NA | 1388131 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |