FDA Adverse Event Other Summary report: N

DA VINCI SURGICAL SYSTEM

MDR report key: 1123768 · Received August 18, 2008

Report

Report Number
2955842-2008-01213
Event Type
Other
Date Received
August 18, 2008
Date of Event
July 22, 2008
Report Date
August 15, 2008
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
NAY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONDUCTED BY FIELD SERVICE ENGINEERING CONFIRMED THAT UNEXPECTED SET UP JOINT MOVEMENT EXPERIENCED BY THE CUSTOMER WAS ASSOCIATED WITH THE ENDOSCOPIC CAMERA MANIPULATOR (ECM). THE ECM IS THE CAMERA ARM LOCATED ON THE PT SIDE CART WHICH PROVIDES THE STERILE INTERFACE FOR THE 3D ENDOSCOPE. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED ECM. THE FIELD SERVICE ENGINEER PERFORMED A CLUTCH MOVE OF THE ECM SET UP JOINT WHICH CLEARED THE SYSTEM ERROR MESSAGE. AS OF 2008, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SURGICAL PROCEDURE, THE CUSTOMER EXPERIENCED UNEXPECTED SET UP JOINT MOVEMENT ON THE ENDOSCOPIC CAMERA MANIPULATOR (ECM) ARM. AN ISI TECHNICAL SUPPORT ENGINEER ATTEMPTED TO ASSIST THE CUSTOMER, HOWEVER, THE SURGEON DECIDED TO CONVERT THE PROCEDURE TO TRADITIONAL OPEN SURGICAL TECHNIQUE TO COMPLETE THE PLANNED PROCEDURE. NO PT HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL, INC. ISI200 A4.3PL

Patients

Seq Age Sex Outcome Treatment
1 Other ACCESSORIES| DA VINCI SURGICAL SYSTEM INSTRUMENTS