FDA Adverse Event Malfunction Summary report: N

SYRINGE S2 10ML 22GA 1-1/4IN BD CHINA

MDR report key: 11234148 · Received January 26, 2021

Report

Report Number
3002682307-2021-00026
Event Type
Malfunction
Date Received
January 26, 2021
Date of Event
January 5, 2021
Report Date
April 26, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 1901142. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE ORIGINALLY FOUND TO BE WITHIN SPECIFICATION. ONE PICTURE SAMPLE WAS RETURNED; HOWEVER, THE PICTURE DID NOT DISPLAY THE PRODUCT RELATED TO THIS REPORTED COMPLAINT. TO FURTHER INVESTIGATE THIS ISSUE, TWENTY RETAINED SAMPLES OF THE SAME LOT NUMBER WERE OBTAINED FROM THE MANUFACTURING FACILITY. THE RETAINED SAMPLES WERE INSPECTED AND NO DEFECTS WERE IDENTIFIED. BASED ON THE INVESTIGATION RESULTS, AN EXACT ISSUE RELATED TO THE MANUFACTURING PROCESS COULD NOT BE DETERMINED.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE S2 10ML 22GA 1-1/4IN BD CHINA LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE USER FOUND LIQUID LEAKAGE AT THE JOINT OF THE SYRINGE PLASTIC AND STEEL NEEDLE DURING THE USE OF THE PRODUCT."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE S2 10ML 22GA 1-1/4IN BD (B)(6) LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE USER FOUND LIQUID LEAKAGE AT THE JOINT OF THE SYRINGE PLASTIC AND STEEL NEEDLE DURING THE USE OF THE PRODUCT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
129832 SYRINGE S2 10ML 22GA 1-1/4IN BD CHINA SYRINGE FMI BECTON DICKINSON, S.A. 1901142

Patients

Seq Age Sex Outcome Treatment
1