FDA Adverse Event Malfunction Summary report: N

SYRINGE S2 2ML

MDR report key: 11233938 · Received January 26, 2021

Report

Report Number
3002682307-2021-00025
Event Type
Malfunction
Date Received
January 26, 2021
Date of Event
January 5, 2021
Report Date
April 27, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 2003184. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE ORIGINALLY FOUND TO BE WITHIN SPECIFICATION. TO AID IN THE INVESTIGATION OF THIS ISSUE, PICTURE SAMPLES WERE RETURNED FOR EVALUATION BY OUR QUALITY ENGINEER TEAM. THROUGH EXAMINATION OF THE PICTURES, LEAKAGE WAS OBSERVED THROUGH THE PLUNGER ROD. IT HAS BEEN DETERMINED THAT THIS INCIDENT WAS A CONSEQUENCE OF DAMAGE TO THE PLUNGER LIP COMPONENT. THIS TYPE OF DAMAGE CAN RESULT DURING THE HANDLING OF THE PRODUCT THROUGH THE MANUFACTURING PROCESS OR WITHIN THE PLUNGER ASSEMBLY MACHINE.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE S2 2ML LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "LEAKAGE PAST STOPPER ON A SYRINGE FILLED WITH METHOTREXATE."

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE S2 2ML LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "LEAKAGE PAST STOPPER ON A SYRINGE FILLED WITH METHOTREXATE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
129682 SYRINGE S2 2ML SYRINGE FMI BECTON DICKINSON, S.A. 2003184

Patients

Seq Age Sex Outcome Treatment
1