INGEVITY+
Report
- Report Number
- 2124215-2021-02271
- Event Type
- Injury
- Date Received
- January 26, 2021
- Date of Event
- October 13, 2020
- Report Date
- June 24, 2021
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- NVN
- UDI-DI
- 00802526604522
- PMA / PMN Number
- P150012/S083
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. TESTING WAS COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INNER INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES. LABORATORY ANALYSIS DID NOT IDENTIFY ANY LEAD CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS.
IT WAS REPORTED THAT THIS PATIENT PRESENTED WITH INCREASING SHORTNESS OF BREATH SINCE IMPLANT IN OCTOBER. THE PHYSICIAN SUSPECTED PACEMAKER INDUCTED CARDIOMYOPATHY. THIS RIGHT VENTRICULAR (RV) LEAD WAS EXPLANTED FROM THE APEX. A NEW LEAD WAS SUCCESSFULLY PLACED IN THE BUNDLE OF HIS TO OPTIMIZE CARDIAC FUNCTION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
(B)(4).
IT WAS REPORTED THAT THIS PATIENT PRESENTED WITH INCREASING SHORTNESS OF BREATH SINCE IMPLANT IN OCTOBER. THE PHYSICIAN SUSPECTED PACEMAKER INDUCTED CARDIOMYOPATHY. THIS RIGHT VENTRICULAR (RV) LEAD WAS EXPLANTED FROM THE APEX. A NEW LEAD WAS SUCCESSFULLY PLACED IN THE BUNDLE OF HIS TO OPTIMIZE CARDIAC FUNCTION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 130525 | INGEVITY+ | IMPLANTABLE LEAD | NVN | BOSTON SCIENTIFIC CORPORATION | 7842 | 1015898 | 00802526604522 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Hospitalization| R |