FDA Adverse Event Injury Summary report: N

INGEVITY+

MDR report key: 11233752 · Received January 26, 2021

Report

Report Number
2124215-2021-02271
Event Type
Injury
Date Received
January 26, 2021
Date of Event
October 13, 2020
Report Date
June 24, 2021
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
NVN
UDI-DI
00802526604522
PMA / PMN Number
P150012/S083
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. TESTING WAS COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INNER INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES. LABORATORY ANALYSIS DID NOT IDENTIFY ANY LEAD CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS PATIENT PRESENTED WITH INCREASING SHORTNESS OF BREATH SINCE IMPLANT IN OCTOBER. THE PHYSICIAN SUSPECTED PACEMAKER INDUCTED CARDIOMYOPATHY. THIS RIGHT VENTRICULAR (RV) LEAD WAS EXPLANTED FROM THE APEX. A NEW LEAD WAS SUCCESSFULLY PLACED IN THE BUNDLE OF HIS TO OPTIMIZE CARDIAC FUNCTION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THIS PATIENT PRESENTED WITH INCREASING SHORTNESS OF BREATH SINCE IMPLANT IN OCTOBER. THE PHYSICIAN SUSPECTED PACEMAKER INDUCTED CARDIOMYOPATHY. THIS RIGHT VENTRICULAR (RV) LEAD WAS EXPLANTED FROM THE APEX. A NEW LEAD WAS SUCCESSFULLY PLACED IN THE BUNDLE OF HIS TO OPTIMIZE CARDIAC FUNCTION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
130525 INGEVITY+ IMPLANTABLE LEAD NVN BOSTON SCIENTIFIC CORPORATION 7842 1015898 00802526604522

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| R