FDA Adverse Event Malfunction Summary report: N

I-STAT1 ANALYZER, IMMUNO READY, WIRELESS

MDR report key: 11231121 · Received January 26, 2021

Report

Report Number
2245578-2021-00006
Event Type
Malfunction
Date Received
January 26, 2021
Date of Event
January 22, 2021
Report Date
April 20, 2021
Manufacturer
ABBOTT POINT OF CARE
Product Code
CGA
UDI-DI
00054749000340
PMA / PMN Number
K001387
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT: 851234 THE INVESTIGATION WAS COMPLETED ON 19-APR-2021. THE CUSTOMER REPORTED THAT "SMOKE BURN" WAS SEEN INSIDE ANALYZER S/N 353367 AND THE BATTERY DOOR WAS OBSERVED BLACK IN COLOR. THE BUSINESS PARTNER ADDED THAT THE CABLES AND/OR BATTERIES IN USE WERE NOT PROVIDED BY APOC. THE CONCLUSION OF THE INVESTIGATION IS THAT THE COMPLAINT WAS CONFIRMED, AND THE LIKELY CAUSE WAS DETERMINED TO BE THE NON-APOC SUPPLIED BATTERIES AND/OR CABLES THAT WERE IN USE AS CONFIRMED BY THE BUSINESS PARTNER. A ROCKETWARE SEARCH SPANNING SIX MONTHS REVEALED NO SIMILAR INCIDENTS AND NO EVIDENCE OF A TREND. NO DEFICIENCY HAS BEEN IDENTIFIED.

Additional Manufacturer Narrative · 1

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 1

ON 22-JAN-2021, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT ANALYZER SN (B)(4) SMOKE BURN WAS SEEN IN THE BATTERY COMPARTMENT AREA OF THE ANALYZER. THERE WAS NO ADDITIONAL INFORMATION AT THE TIME OF THIS REPORT. THE ANALYZER WAS REPLACED AT NO CHARGE AND RETURNING FOR INVESTIGATION. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. CARRIER INFORMATION WAS NOT AVAILABLE. HOWEVER, THE ANALYZER SHIPPED ON 07-DEB-20212 WITH DISPOSABLE ALKALINE 9V DURACELL BATTERIES AND A RED CARRIER. THERE IS NO REASON TO SUSPECT THE PRODUCT WOULD BECOME HOT TO TOUCH. THE PRODUCT WAS REPLACED AND RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
126714 I-STAT1 ANALYZER, IMMUNO READY, WIRELESS I-STAT1 ANALYZER CGA ABBOTT POINT OF CARE NA NA 00054749000340

Patients

Seq Age Sex Outcome Treatment
1