ENERGEN ICD
Report
- Report Number
- 2124215-2020-29012
- Event Type
- Malfunction
- Date Received
- January 25, 2021
- Date of Event
- May 2, 2020
- Report Date
- May 21, 2022
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- LWP
- UDI-DI
- 00802526480744
- PMA / PMN Number
- P960040/S235
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS PRODUCT HAS NOT BEEN RETURNED TO BOSTON SCIENTIFIC, AND AS A RESULT, LABORATORY ANALYSIS COULD NOT BE CONDUCTED. THE ASSOCIATED INVESTIGATION DETERMINED THAT THERE WAS NO CONCLUSIVE EVIDENCE THAT THE OBSERVED OUT-OF-RANGE IMPEDANCE MEASUREMENTS WERE DUE TO A MALFUNCTION OR AN INADEQUATE LEAD-TO-DEVICE CONNECTION. HOWEVER, INTERMITTENT, OUT-OF-RANGE SHOCK LEAD IMPEDANCE MEASUREMENTS WITH NO CONCLUSIVE EVIDENCE OF A MALFUNCTION OR AN INADEQUATE LEAD-TO-DEVICE CONNECTION ARE LIKELY THE RESULT OF THE LOW-ENERGY TEST SIGNAL UTILIZED FOR DAILY AUTOMATIC OR IN-CLINIC COMMANDED SHOCK LEAD IMPEDANCE TESTING. A SOFTWARE UPDATE WAS RELEASED IN 2015 TO FURTHER IMPROVE CONSISTENCY OF THE IMPEDANCE TEST RESULTS AND PROVIDE THE CLINICIAN WITH ADDITIONAL DIAGNOSTIC TOOLS, INCLUDING PROGRAMMABLE SHOCK LEAD IMPEDANCE ALERT LIMITS. -- THE RETURNED DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY. VISUAL EXAMINATION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. THE ASSOCIATED INVESTIGATION DETERMINED THAT THERE WAS NO CONCLUSIVE EVIDENCE THAT THE OBSERVED OUT-OF-RANGE IMPEDANCE MEASUREMENTS WERE DUE TO A MALFUNCTION OR AN INADEQUATE LEAD-TO-DEVICE CONNECTION. HOWEVER, INTERMITTENT, OUT-OF-RANGE SHOCK LEAD IMPEDANCE MEASUREMENTS WITH NO CONCLUSIVE EVIDENCE OF A MALFUNCTION OR AN INADEQUATE LEAD-TO-DEVICE CONNECTION ARE LIKELY THE RESULT OF THE LOW-ENERGY TEST SIGNAL UTILIZED FOR DAILY AUTOMATIC OR IN-CLINIC COMMANDED SHOCK LEAD IMPEDANCE TESTING. A SOFTWARE UPDATE WAS RELEASED IN 2015 TO FURTHER IMPROVE CONSISTENCY OF THE IMPEDANCE TEST RESULTS AND PROVIDE THE CLINICIAN WITH ADDITIONAL DIAGNOSTIC TOOLS, INCLUDING PROGRAMMABLE SHOCK LEAD IMPEDANCE ALERT LIMITS.
IT WAS REPORTED THAT THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND DEFIBRILLATION LEAD EXHIBITED HIGH OUT-OF-RANGE SHOCK IMPEDANCE MEASUREMENTS GREATER THAN 125 OHMS. IT WAS NOTED THAT THE DEVICE HAD NOT DELIVERED SHOCKS SINCE IMPLANT (2013).TECHNICAL SERVICES (TS) DISCUSSED POSSIBLE CAUSES INCLUDING CALCIFICATION. TS ALSO DISCUSSED INCREASING THE ALERT LIMIT AND CONSIDERING COMMANDED SHOCKS. THIS DEVICE AND LEAD REMAIN IN SERVICE AT THIS TIME. NO ADVERSE PATIENT EFFECTS WERE REPORTED. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED AND UPGRADED TO A CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM. HOWEVER THE RIGHT VENTRICULAR (RV) LEAD REMAINS IN SERVICE. IT WAS NOTED THAT SHOCK IMPEDANCE MEASUREMENT AT THE UPGRADE PROCEDURE WAS 42 OHMS. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS ICD WAS RETURNED FOR ANALYSIS.
THIS PRODUCT HAS NOT BEEN RETURNED TO BOSTON SCIENTIFIC, AND AS A RESULT, LABORATORY ANALYSIS COULD NOT BE CONDUCTED. THE ASSOCIATED INVESTIGATION DETERMINED THAT THERE WAS NO CONCLUSIVE EVIDENCE THAT THE OBSERVED OUT-OF-RANGE IMPEDANCE MEASUREMENTS WERE DUE TO A MALFUNCTION OR AN INADEQUATE LEAD-TO-DEVICE CONNECTION. HOWEVER, INTERMITTENT, OUT-OF-RANGE SHOCK LEAD IMPEDANCE MEASUREMENTS WITH NO CONCLUSIVE EVIDENCE OF A MALFUNCTION OR AN INADEQUATE LEAD-TO-DEVICE CONNECTION ARE LIKELY THE RESULT OF THE LOW-ENERGY TEST SIGNAL UTILIZED FOR DAILY AUTOMATIC OR IN-CLINIC COMMANDED SHOCK LEAD IMPEDANCE TESTING. A SOFTWARE UPDATE WAS RELEASED IN 2015 TO FURTHER IMPROVE CONSISTENCY OF THE IMPEDANCE TEST RESULTS AND PROVIDE THE CLINICIAN WITH ADDITIONAL DIAGNOSTIC TOOLS, INCLUDING PROGRAMMABLE SHOCK LEAD IMPEDANCE ALERT LIMITS.
IT WAS REPORTED THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND DEFIBRILLATION LEAD EXHIBITED HIGH OUT-OF-RANGE SHOCK IMPEDANCE MEASUREMENTS GREATER THAN 125 OHMS. IT WAS NOTED THAT THE DEVICE HAD NOT DELIVERED SHOCKS SINCE IMPLANT (2013).TECHNICAL SERVICES (TS) DISCUSSED POSSIBLE CAUSES INCLUDING CALCIFICATION. TS ALSO DISCUSSED INCREASING THE ALERT LIMIT AND CONSIDERING COMMANDED SHOCKS. THIS DEVICE AND LEAD REMAIN IN SERVICE AT THIS TIME. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 121316 | ENERGEN ICD | IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) | LWP | BOSTON SCIENTIFIC CORPORATION | E143 | 108005 | 00802526480744 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Male |