ENDOPATH ENDOSCOPIC MULTIFEED STAPLER
Report
- Report Number
- 1527736-1997-01921
- Event Type
- Malfunction
- Date Received
- August 7, 1997
- Report Date
- July 11, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GAG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION. THE RESULTS OF THE INVESTIGATION CONDUCTED BY THE APPROPRIATE ENGINEERS AND TECHNICIANS ARE LISTED BELOW. VISUAL INSPECTIONS & RESULTS: HEAD ROTATES, ABC)YES; NOSE SHROUD CRACKED/BROKEN, AC)NO,B)YES; STAPLES IN NOSE, AC)NO,B)YES(5); STAPLES IN THE TRACK, AC(0),B(11) AND TRIGGER ENGAGED WITH PRECOCK, ABC) YES; FUNCTIONAL TESTS & RESULTS: CYCLED INSTRUMENT, AC)NO,B)YES. ANALYSIS CONCLUSION: BASED UPON THE INQUIRY INFO REC'D, VISUAL EXAMINATION AND FUNCTIONAL TEST, IT WAS CONCLUDED THAT THE REPORTED "DEVICES JAMMED" DURING SURGERY OCCURRED POSSIBLY DUE TO FIVE STAPLES JAMMED IN THE NOSE AND A YIELDED CARTRIDGE NOSE WELD FROM INSTRUMENT (B). THE DEVICE FIRED TWO STAPLES WELL FORMED FOLLOWED BY FIVE MALFORMED STAPLES DUE TO THE YIELDED NOSE WELD. NO CONCLUSION COULD BE REACHED WHETHER THE FIVE STAPLES JAMMED IN THE NOSE CAUSED THE NOSE WELD TO YIELD, OR IF THE YIELDED NOSE WELD CAUSED THE FIVE STAPLES TO BECOME JAMMED IN THE NOSE. INSTRUMENTS (A) AND (C) WERE REC'D EMPTY IN GOOD PHYSICAL CONDITION. CO REVIEWS EACH INCIDENT AS IT OCCURS IN AN EFFORT TO CONTINUOUSLY IMPROVE CO'S PRODUCTS AND PROCESSES.
IT WAS REPORTED THE DEVICES WERE USED DURING AN UNK PROCEDURE. IT WAS REPORTED BY THE AFFILIATE THAT THE DEVICES JAMMED. IT WAS NOT REPORTED IF THERE WERE ANY PT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH ENDOSCOPIC MULTIFEED STAPLER | ENDOSCOPIC STAPLER | GAG | ETHICON ENDO-SURGERY, INC. | NA | J4526Z |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |