FDA Adverse Event Malfunction Summary report: N

ENDOPATH ENDOSCOPIC MULTIFEED STAPLER

MDR report key: 112234 · Received August 7, 1997

Report

Report Number
1527736-1997-01931
Event Type
Malfunction
Date Received
August 7, 1997
Report Date
July 11, 1997
Manufacturer
ETHICON ENDO-SURGERY, INC.
Product Code
GAG
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

FACILITY EXPERIENCED AN EVENT WITH ENDOPATH* ENDOSCOPIC MULTIFEED STAPLER WHILE PERFORMING A UNKNOWN PROCEDURE. THE PRODUCT COMPLAINT ANALYSIS TEAM HAD COMPLETED ITS INVESTIGATION OF THE DEVICE WHICH WAS RETURNED TO CO WITH PRODUCT INQUIRY #974387. THE RESULTS OF THE INVESTIGATION CONDUCTED BY THE APPROPRIATE ENGINEERS AND TECHNICIANS ARE LISTED BELOW. VISUAL INSPECTIONS & RESULTS: STAPLES IN NOSE, YES(1 TWISTED). STAPLES IN THE TRACK, TRIGGER ENGAGED WITH PRECOCK, FUNCTIONAL TESTS & RESULTS: CYLCED INSTRUMENT, NO. ANALYSIS CONCLUSION: BASED UPON THE INQUIRY INFO RECEIVED AND THE VISUAL EXAMINATION, NO CONCLUSION COULD BE REACHED WHY IT WAS REPORTED "STAPLES WOULD NOT ADVANCE" DURING SURGERY. THE INSTRUMENT WAS RECEIVED WITH A TWISTED STAPLE IN THE NOSE AND A SEVERELY BENT TUBE (WHICH OCCURRED DURING SURGERY). NO FUNCTIONAL ANALYSIS COULD BE PERFORMED DUE TO THE BENT TUBE. CO REVIEWS EACH INCIDENT AS IT OCCURS IN AN EFFORT TO CONTINUOUSLY IMPROVE THE PRODUCTS AND PROCESSES.

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE WAS USED DURING AN UNK PROCEDURE. IT WAS REPORTED BY THE AFFILIATE THAT THE STAPLES WOULD NOT ADVANCE. THERE WAS NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ENDOSCOPIC MULTIFEED STAPLER ENDOSCOPIC STAPLER GAG ETHICON ENDO-SURGERY, INC. NA K46L47

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other