FDA Adverse Event
Malfunction
Summary report: N
CENTURY 835 BED
MDR report key: 1122221
·
Received August 15, 2008
Report
- Report Number
- 1824206-2008-03133
- Event Type
- Malfunction
- Date Received
- August 15, 2008
- Date of Event
- July 17, 2008
- Report Date
- July 17, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HEAD SECTION DRIFTING IS UNINTENTIONAL MOVEMENT OF THE BED WHICH HAS THE POTENTIAL TO CAUSE SERIOUS INJURY. THE TECHNICIAN CLEANED AND LUBRICATED THE HEAD DRIVE IN ORDER TO RESOLVE THE PROBLEM. ALL BED DRIVES ARE REQUIRED TO BE INSPECTED, CLEANED, AND LUBRICATED AT LEAST ONCE A YEAR IN ACCORDANCE WITH THE PRODUCT SERVICE MANUAL.
Description of Event or Problem · 1
AN ACCOUNT REPORTED THAT THE HEAD SECTION ON THIS BED WOULD DRIFT DOWN AFTER BEING RAISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CENTURY 835 BED | AC POWERED HOSPITAL BED | FNL | HILL-ROM RITTER | P835 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |