FDA Adverse Event Malfunction Summary report: N

CENTURY 835 BED

MDR report key: 1122221 · Received August 15, 2008

Report

Report Number
1824206-2008-03133
Event Type
Malfunction
Date Received
August 15, 2008
Date of Event
July 17, 2008
Report Date
July 17, 2008
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HEAD SECTION DRIFTING IS UNINTENTIONAL MOVEMENT OF THE BED WHICH HAS THE POTENTIAL TO CAUSE SERIOUS INJURY. THE TECHNICIAN CLEANED AND LUBRICATED THE HEAD DRIVE IN ORDER TO RESOLVE THE PROBLEM. ALL BED DRIVES ARE REQUIRED TO BE INSPECTED, CLEANED, AND LUBRICATED AT LEAST ONCE A YEAR IN ACCORDANCE WITH THE PRODUCT SERVICE MANUAL.

Description of Event or Problem · 1

AN ACCOUNT REPORTED THAT THE HEAD SECTION ON THIS BED WOULD DRIFT DOWN AFTER BEING RAISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTURY 835 BED AC POWERED HOSPITAL BED FNL HILL-ROM RITTER P835 NA

Patients

Seq Age Sex Outcome Treatment
1