FDA Adverse Event Malfunction Summary report: N

4 MOTOR ADVANCE BED

MDR report key: 1122218 · Received August 15, 2008

Report

Report Number
1824206-2008-03132
Event Type
Malfunction
Date Received
August 15, 2008
Date of Event
July 15, 2008
Report Date
July 15, 2008
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE HOSPITAL MAINTENANCE PERSON REPORTED THAT THE HEAD FUNCTION RAN UP UNINTENTIONALLY. HE FOUND THAT THE HEAD UP PATIENT SWITCH WAS STUCK IN THE LEFT HEAD SIDERAIL. HE REPLACED THE SWITCH ASSEMBLY AND THE BED FUNCTIONED AS DESIGNED. HE STATED THAT THERE WERE NO INJURIES ASSOCIATED WITH THIS PRODUCT MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 4 MOTOR ADVANCE BED AC POWERED HOSPITAL BED FNL HILL-ROM RITTER 1105 NA

Patients

Seq Age Sex Outcome Treatment
1