LIGAMAX-5MM ENDO CLIP APPLIER
Report
- Report Number
- 3005075853-2008-01217
- Event Type
- Malfunction
- Date Received
- August 14, 2008
- Date of Event
- July 9, 2008
- Report Date
- July 11, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THE EL5ML DEVICE WAS REC'D IN GOOD VISUAL CONDITION. WHEN TESTING THE DEVICE FOR FUNCTIONALITY, IT WAS NOTED TO BE EMPTY AND THE LOCKOUT WAS FIRED THROUGH. AS THE IFU STATES: "DO NOT ATTEMPT TO FIRE THROUGH THE LOCKOUT. IF THE TRIGGER IS FORCED CLOSED ONCE THE LAST CLIP LOCKOUT HAS ENGAGED, THE JAWS MAY REMAIN CLOSED". WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A GASTRECTOMY PROCEDURE, THE DEVICE WOULD NOT CLIP AT THE FIFTH AND SIXTH FIRINGS WITH A CRASHING NOISE. ALTHOUGH THE DEVICE WAS NOT USED MANY TIMES, THE INDICATOR WAS ALREADY SHOWN. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGAMAX-5MM ENDO CLIP APPLIER | FZP | ETHICON ENDO-SURGERY, LLC | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |