FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 1122036 · Received August 14, 2008

Report

Report Number
3005075853-2008-01217
Event Type
Malfunction
Date Received
August 14, 2008
Date of Event
July 9, 2008
Report Date
July 11, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THE EL5ML DEVICE WAS REC'D IN GOOD VISUAL CONDITION. WHEN TESTING THE DEVICE FOR FUNCTIONALITY, IT WAS NOTED TO BE EMPTY AND THE LOCKOUT WAS FIRED THROUGH. AS THE IFU STATES: "DO NOT ATTEMPT TO FIRE THROUGH THE LOCKOUT. IF THE TRIGGER IS FORCED CLOSED ONCE THE LAST CLIP LOCKOUT HAS ENGAGED, THE JAWS MAY REMAIN CLOSED". WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A GASTRECTOMY PROCEDURE, THE DEVICE WOULD NOT CLIP AT THE FIFTH AND SIXTH FIRINGS WITH A CRASHING NOISE. ALTHOUGH THE DEVICE WAS NOT USED MANY TIMES, THE INDICATOR WAS ALREADY SHOWN. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM ENDO CLIP APPLIER FZP ETHICON ENDO-SURGERY, LLC NA UNK

Patients

Seq Age Sex Outcome Treatment
1