FDA Adverse Event Malfunction Summary report: N

*

MDR report key: 1121950 · Received June 12, 2008

Report

Report Number
2250051-2008-00284
Event Type
Malfunction
Date Received
June 12, 2008
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS, INC.
Product Code
JTC
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SVC ENGINEER (FSE) EVALUATED THE INSTRUMENT AND FOUND A BROKEN CLAMP IN THE REPORTED PROBLEM AREA OF THE PIPETTE ASSEMBLY. REPLACED PLUNGER CLAMP IN POSITION #7. THE INSTRUMENT WAS CLEANED, TESTED WITHOUT FURTHER PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 * * JTC ORTHO-CLINICAL DIAGNOSTICS, INC. * *

Patients

Seq Age Sex Outcome Treatment
1