FDA Adverse Event
Malfunction
Summary report: N
ULTRACISION HARMONIC SCALPEL HAND PIECE
MDR report key: 1121909
·
Received August 13, 2008
Report
- Report Number
- 3005075853-2008-01167
- Event Type
- Malfunction
- Date Received
- August 13, 2008
- Date of Event
- July 8, 2008
- Report Date
- July 9, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- LFL
- PMA / PMN Number
- K002906
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: THE HAND PIECE WAS RETURNED WITH A LOOSE MOUNT AND THE NOSE CONE CRACKED. IT WAS TESTED ON A GENERATOR AND NO ERROR CODE WAS NOTED; IT ACTIVATED PROPERLY. HOWEVER, THE HAND PIECE NO LONGER MEETS THE DESIGN SPECIFICATIONS FOR CONTINUITY. THE HAND PIECE WAS DISASSEMBLED AND A TORN ACOUSTIC ISOLATOR WAS FOUND. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A GYN PROCEDURE, THE DEVICE WOULD NOT ACTIVATE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRACISION HARMONIC SCALPEL HAND PIECE | LFL | ETHICON ENDO-SURGERY, LLC | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HANDPIECE| GENERATOR |