FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL HAND PIECE

MDR report key: 1121909 · Received August 13, 2008

Report

Report Number
3005075853-2008-01167
Event Type
Malfunction
Date Received
August 13, 2008
Date of Event
July 8, 2008
Report Date
July 9, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
LFL
PMA / PMN Number
K002906
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE HAND PIECE WAS RETURNED WITH A LOOSE MOUNT AND THE NOSE CONE CRACKED. IT WAS TESTED ON A GENERATOR AND NO ERROR CODE WAS NOTED; IT ACTIVATED PROPERLY. HOWEVER, THE HAND PIECE NO LONGER MEETS THE DESIGN SPECIFICATIONS FOR CONTINUITY. THE HAND PIECE WAS DISASSEMBLED AND A TORN ACOUSTIC ISOLATOR WAS FOUND. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A GYN PROCEDURE, THE DEVICE WOULD NOT ACTIVATE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL HAND PIECE LFL ETHICON ENDO-SURGERY, LLC NA NA

Patients

Seq Age Sex Outcome Treatment
1 HANDPIECE| GENERATOR