8015 ALARIS SYSTEM PC UNIT
Report
- Report Number
- 2016493-2021-14885
- Event Type
- Malfunction
- Date Received
- January 22, 2021
- Report Date
- April 17, 2020
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403801518
- PMA / PMN Number
- K133532
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THE CUSTOMER¿S REPORTED PROBLEM WAS CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS NOT RETURNED FOR SERVICING WHICH NOT CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
(B)(4). PATIENT OR USER INVOLVEMENT: NO. PATIENT OR USER HARM: NO. CASE DESCRIPTION: TECH CALLED IN FOR HELP ON ERROR CODE ON 8015 BD UNIT SERIAL # (B)(4). CASE RESOLUTION: TECH IS GET ERROR CODE 810.9030 ON 8015 BD UNIT BEFORE CALL IN TECH TRY TO REFLASH SOFTWARE FAIL, EXPLAIN TO TECH YES THAT IS THE FIRST STEP TO DO BECAUSE ITS THE POLO IS NOT ABLE TO BOOT BECAUSE THE APPLICATION ELF FILE HAS AN INVALID CHECKSUM. , NEXT WOULD BE TO REPLACE IS THE MAIN BOARD ON THE UNIT, TECH PREFER TO SEND UNIT IN FOR REPAIR CONFIRM PRICE QUOTE FOR LEVEL ONE REPAIR. TECH WILL CALL BACK WHEN HE HAS HIS PO# TO SEND UNIT IN FOR REPAIR SERVICES. TECH THANK ME FOR MY ASSIST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 113042 | 8015 ALARIS SYSTEM PC UNIT | PUMP, INFUSION | FRN | CAREFUSION SD | 8015 | 10885403801518 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |