LIFEVEST WCD 3000 SYSTEM
Report
- Report Number
- 3002158293-2008-00365
- Event Type
- Malfunction
- Date Received
- August 14, 2008
- Date of Event
- May 15, 2008
- Report Date
- August 13, 2008
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY PACK HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE CAUSE OF THE FAULTS WAS DEFECTIVE BATTERY CELLS WITHIN THE BATTERY PACK. THE ROOT CAUSE OF THE DEFECTIVE CELLS IS NOT KNOWN, BUT WAS PROBABLY DUE TO EXCESSIVE DISCHARGING. MANY "BATTERY RUNTIME EXPIRED" FLAGS WERE SEEN ON THE FINAL DOWNLOAD FROM THE LAST PATIENT TO USE THIS BATTERY PACK. THIS MEANT THAT THE BATTERY PACK WAS USED FOR GREATER THAN TWENTY-FOUR HOURS. THIS MEANS THAT THE PATIENT CONTINUED TO USE A DEPLETED BATTERY FOR HOURS AFTER THE EXPIRED RUNTIME ALARMS SOUNDED. THE DEFECTIVE CELLS WERE REPLACED. THE BATTERY PACK WAS RETESTED AND RESTOCKED. NO ADVERSE EVENT RESULTED FROM THE FAULTY BATTERY PACK. THE PATIENT RECEIVED A REPLACEMENT BATTERY PACK.
A FOREIGN DISTRIBUTOR SENT AN EMAIL CONTAINING A SERVICE REQUEST FOR A BATTERY PACK. RESEARCH REVEALED THE LAST PATIENT TO USE THIS BATTERY PACK WAS EXCESSIVELY DISCHARGING THE PACK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |