FDA Adverse Event
Malfunction
Summary report: N
CONTOUR TEST STRIPS (50)
MDR report key: 1120983
·
Received August 13, 2008
Report
- Report Number
- 1826988-2008-00867
- Event Type
- Malfunction
- Date Received
- August 13, 2008
- Date of Event
- July 16, 2008
- Report Date
- July 16, 2008
- Manufacturer
- BAYER HEALTHCARE, LLC
- Product Code
- NBW
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER ALLEGED THAT SHE WAS RECEIVING HIGH GLUCOSE READINGS USING HER CONTOUR METER. WHILE TROUBLESHOOTING, SHE PERFORMED CONTROL TESTS AND RECEIVED A RESULT OF 249 MG/DL. THE NORMAL CONTROL RANGE WAS 102-141 MG/DL. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS WERE RETURNED FOR EVAL. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTOUR TEST STRIPS (50) | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE, LLC | NOT PROVIDED | 7HC3D07 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |