FDA Adverse Event Injury Summary report: N

HEALTHSMART

MDR report key: 11207444 · Received January 21, 2021

Report

Report Number
1422443-2021-00001
Event Type
Injury
Date Received
January 21, 2021
Date of Event
January 3, 2021
Report Date
January 21, 2021
Manufacturer
CARDICARE COMPANY LTD
Product Code
IRT
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INSPECTED UNIT. UNIT LOOKS TO HAVE BEEN USED EXCESSIVELY. HEATING PAD IS TORN WHERE THE ELECTRICAL CABLE COMES OUT AT THE TOP RIGHT CORNER. MIGHT BE DUE TO MISHANDLING OF PRODUCT. PULL AND/OR TUGGING OF UNIT. PLUGGED IN UNIT INTO OUTLET. UNIT SUCCESSFULLY POWERS ON. CYCLED THROUGH ALL HEAT LEVELS OF THE UNIT TO VERIFY ALL COLOR INDICATOR LAMPS WERE FULLY FUNCTIONAL. TURNED ON UNIT TO HIGH SETTING AT 9:10AM. VERIFIED UNIT WAS HEATING AND STILL POWERED ON EVERY 15 MINUTES UNTIL UNIT AUTOMATICALLY POWERED OFF AT 11:10AM. TURNED ON UNIT TO MEDIUM SETTING AT 11:30AM. VERIFIED UNIT WAS HEATING AND STILL POWERED ON EVERY 15 MINUTES UNTIL UNIT AUTOMATICALLY POWERED OFF AT 1:30PM. UNIT SEEMS TO BE FUNCTIONING NORMAL.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE END USER DOZED OFF WHILE USING THE HEATING PAD AND RECEIVED A BURNED ON HER BACK. THE END USER CONTACTED HER DOCTOR AND HAS BEEN APPLYING HYDROCORTISONE OINTMENT WITH PAIN RELIEVER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE END USER DOZED OFF WHILE USING THE HEATING PAD AND RECEIVED A BURNED ON HER BACK. THE END USER CONTACTED HER DOCTOR AND HAS BEEN APPLYING HYDROCORTISONE OINTMENT WITH PAINT RELEIVER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
107722 HEALTHSMART HEATING PAD IRT CARDICARE COMPANY LTD 619-5133-1900 90241218

Patients

Seq Age Sex Outcome Treatment
1 40 YR Other