DA VINCI S SURGICAL SYSTEM
Report
- Report Number
- 2955842-2008-01166
- Event Type
- Other
- Date Received
- August 14, 2008
- Date of Event
- May 9, 2008
- Report Date
- August 14, 2008
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- NAY
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION CONDUCTED BY FIELD SERVICE ENGINEERING CONCLUDED THAT SYSTEM ERROR CODE #212 EXPERIENCED BY THE CUSTOMER WAS ASSOCIATED WITH THE LEFT MASTER TOOL MANIPULATOR (MTML) AND/OR A MULTIPLE SERVO DRIVER (MSD) PCA BOARD. THE MASTER TOOL MANIPULATOR REFERS TO THE MASTER CONTROLLERS WHICH PROVIDE THE MEANS FOR THE SURGEON TO CONTROL THE INSTRUMENTS AND ENDOSCOPE INSIDE THE PT FROM THE SURGEON'S SIDE CONSOLE. ONE MTM IS ASSIGNED TO THE SURGEON'S LEFT HAND (MTML) AND ONE TO THEIR RIGHT (MTMR). THE SYSTEM CONTAINS FIVE MULTIPLE SERVO DRIVER (MSD) BOARDS WHICH PROVIDE DRIVE CURRENT TO THE SERVO MOTORS ASSOCIATED WITH EACH DEGREE OF FREEDOM. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED MTML AND MSD PCA. THE SYSTEM ALARM (SYSTEM GENERATED FAULT CODE) FUNCTIONED AS DESIGNED AND THERE WAS NO INJURY TO THE PT. THE #212 SYSTEM ERROR CODE APPEARS WHEN THE DAVINCI S SAFETY SYSTEM DETECTS AN ERROR IN THE EXPECTED MOTOR CURRENT AND VOLTAGE VALUES. UPON DETERMINING THIS CONDITION, THE SAFETY SYSTEMS PUT DAVINCI S IN A "RECOVERABLE SAFE STATE". THE MTML AND MSD PCA WERE RETURNED FOR EVAL. NO ISSUES WERE FOUND WITH THE MSD PCA, HOWEVER, ENGINEERING WAS ABLE TO REPLICATE SYSTEM ERROR CODE #212 DURING TESTING OF THE MTML. FURTHER EVAL OF THE MTML REVEALED THAT AN ELECTRICAL CABLE/HARNESS HAD MALFUNCTIONED, THUS GENERATING THE SYSTEM ERROR CODE EXPERIENCED BY THE CUSTOMER. AS OF 2008, THERE HAVE BEEN NO REPORTED RECURRENCE OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED THAT DURING A DA VINCI S PARTIAL NEPHRECTOMY SURGICAL PROCEDURE, THE CUSTOMER EXPERIENCED NONRECOVERABLE SYSTEM ERROR CODE #212. AN ISI TECHNICAL SUPPORT ENGINEER ATTEMPTED TO TROUBLESHOOT THE ISSUE VIA TELEPHONE, HOWEVER, THE ISSUE COULD NOT BE RESOLVED AND THE PLANNED SURGICAL PROCEDURE WAS ABORTED AND RESCHEDULED FOR A LATER DATE. NO PT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI S SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL, INC. | IS2000 A5. 1P4 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |