FDA Adverse Event Malfunction Summary report: N

LIFEPAK 20 DEFIBRILLATOR/MONITOR

MDR report key: 1120603 · Received August 12, 2008

Report

Report Number
3015876-2008-00933
Event Type
Malfunction
Date Received
August 12, 2008
Date of Event
July 14, 2008
Report Date
July 14, 2008
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
LDD
PMA / PMN Number
K012274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE, AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE POWER SUPPLY ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. PHYSIO-CONTROL FURTHER EVALUATED THE REPLACED ASSEMBLY, AND DETERMINED THAT THE ROOT CAUSE OF THE REPORTED FAILURE WAS AN ELECTRICALLY LEAKY FILTER, DESIGNATOR FL4.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD NOT OPERATE ON BATTERY POWER. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 20 DEFIBRILLATOR/MONITOR LDD PHYSIO-CONTROL, INC. 20 NA

Patients

Seq Age Sex Outcome Treatment
1 NA