FDA Adverse Event Malfunction Summary report: N

LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES

MDR report key: 1120564 · Received August 12, 2008

Report

Report Number
3015876-2008-00938
Event Type
Malfunction
Date Received
August 12, 2008
Date of Event
July 16, 2008
Report Date
July 16, 2008
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K991910
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED INTERMITTENT FAILURE. PHYSIO REPLACED THE MAIN CONTROL KEYPAD ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD INTERMITTENTLY NOT POWER ON. THERE WAS NO PT USE ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 12 DEFIBRILLATOR/MONITOR SERIES MKJ PHYSIO-CONTROL, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA