ARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM, ARCTICGEL PADS
Report
- Report Number
- 1018233-2021-00031
- Event Type
- Malfunction
- Date Received
- January 20, 2021
- Date of Event
- December 31, 2020
- Report Date
- July 16, 2021
- Manufacturer
- MEDIVANCE, INC. ¿ 1725056
- Product Code
- DWJ
- PMA / PMN Number
- K142702
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- PHYSICIAN
Narratives
THE REPORTED EVENT WAS CONFIRMED HOWEVER THE CAUSE WAS UNKNOWN. VISUAL INSPECTION NOTED ONE NEONATAL ARTIC GEL PAD WAS RECEIVED. VISUAL EVALUATION NOTED NO OBVIOUS DEFECTS. THE PAD WAS TESTED AND FOUND THAT A TOTAL OF 0 LPM OF FLOW RATE WERE REGISTERED DURING THE TEST. ALSO THE SYSTEM DIAGNOSTICS WERE REVIEWED AND THE CIRCULATION PUMP COMMAND WAS 46 PERCENT THE INLET PRESSURE WAS -7.1PSI. BASED ON THE TEST METHOD THE FLOW RATE WAS FOUND TO BE UNACCEPTABLE. THE PRODUCT WAS USED FOR TREATMENT PURPOSES. THE PRODUCT DID NOT MEET THE SPECIFICATIONS AND WAS INFLUENCED BY THE REPORTED FAILURE. A POTENTIAL ROOT CAUSE FOR THIS FAILURE MODE COULD BE LOW FLOW OR RESTRICTED DUE TO OVERHEATING OF MATERIALS DURING THE LAMINATION PROCESS AND THE WATER FLOW PATH WAS COMPROMISED. THE DEVICE HISTORY RECORD WAS REVIEWED AND FOUND NOTHING THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE INSTRUCTIONS FOR USE WERE FOUND ADEQUATE AND STATE THE FOLLOWING: "INDICATIONS FOR USE: ¿ THE ARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM IS INTENDED FOR MONITORING AND CONTROLLING PATIENT TEMPERATURE IN ADULT AND PEDIATRIC PATIENTS OF ALL AGES. CONTRAINDICATIONS: ¿ THERE ARE NO KNOWN CONTRAINDICATIONS FOR THE USE OF A THERMOREGULATORY SYSTEM. ¿ DO NOT PLACE ARCTIC GEL PADS ON SKIN THAT HAS SIGNS OF ULCERATIONS, BURNS, HIVES OR RASH. ¿WHILE THERE ARE NO KNOWN ALLERGIES TO HYDROGEL MATERIALS, CAUTION SHOULD BE EXERCISED WITH ANY PATIENT WITH A HISTORY OF SKIN ALLERGIES OR SENSITIVITIES. WARNING: ¿ DO NOT PLACE ARCTIC GEL PADS OVER TRANSDERMAL MEDICATION PATCHES AS WARMING CAN INCREASE DRUG DELIVERY, RESULTING IN POSSIBLE HARM TO THE PATIENT. CAUTIONS: ¿ FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN. ¿ THIS PRODUCT IS TO BE USED BY OR UNDER THE SUPERVISION OF TRAINED, QUALIFIED MEDICAL PERSONNEL. ¿ THE CLINICIAN IS RESPONSIBLE FOR DETERMINING THE APPROPRIATENESS OF USE OF THIS DEVICE AND THE USER SETTABLE PARAMETERS, INCLUDING WATER TEMPERATURE, FOR EACH PATIENT. ¿ DUE TO UNDERLYING MEDICAL OR PHYSIOLOGICAL CONDITIONS, SOME PATIENTS ARE MORE SUSCEPTIBLE TO SKIN DAMAGE FROM PRESSURE AND HEAT OR COLD. PATIENTS AT RISK INCLUDE THOSE WITH POOR TISSUE PERFUSION OR POOR SKIN INTEGRITY DUE TO EDEMA, DIABETES, PERIPHERAL VASCULAR DISEASE, POOR NUTRITIONAL STATUS OR STEROID OR HIGH DOSE VASOPRESSOR THERAPY. IF ACCESSIBLE, EXAMINE THE PATIENTS SKIN UNDER THE ARCTIC GEL PADS OFTEN ESPECIALLY THOSE PATIENTS AT HIGHER RISK OF SKIN INJURY. ¿ SKIN INJURY MAY OCCUR AS A CUMULATIVE RESULT OF PRESSURE, TIME AND TEMPERATURE. POSSIBLE SKIN INJURIES INCLUDE BRUISING, TEARING, SKIN ULCERATIONS, BLISTERING, AND NECROSIS. DO NOT PLACE BEAN BAGS OR OTHER FIRM POSITIONING DEVICES UNDER THE ARCTIC GEL PADS. DO NOT PLACE ANY POSITIONING DEVICES UNDER THE PAD MANIFOLDS OR PATIENT LINES. ¿ DO NOT ALLOW URINE, ANTIBACTERIAL SOLUTIONS OR OTHER AGENTS TO POOL UNDERNEATH THE ARCTIC GEL PADS. URINE AND ANTIBACTERIAL AGENTS CAN ABSORB INTO THE PAD HYDROGEL AND CAUSE CHEMICAL INJURY AND LOSS OF PAD ADHESION. REPLACE PADS IMMEDIATELY IF THESE FLUIDS COME INTO CONTACT WITH THE HYDROGEL. ¿ DO NOT PLACE ARCTIC GEL PADS DIRECTLY OVER AN ELECTROSURGICAL GROUNDING PAD. THE COMBINATION OF HEAT SOURCES MAY RESULT IN SKIN BURNS. ¿ CAREFULLY REMOVE ARCTIC GEL PADS FROM THE PATIENT¿S SKIN AT THE COMPLETION OF USE. AGGRESSIVE REMOVAL OR REMOVAL OF COLD PADS FROM THE PATIENTS SKIN MAY RESULT IN SKIN TEARS. ¿ THE ARCTIC GEL PADS ARE NON STERILE FOR SINGLE PATIENT USE ONLY. DO NOT PLACE PADS IN THE STERILE FIELD. IF USED IN A STERILE ENVIRONMENT, PADS SHOULD BE PLACED ACCORDING TO THE PHYSICIANS DIRECTIONS, EITHER PRIOR TO THE STERILE PREPARATION OR STERILE DRAPING. ¿ DO NOT REPROCESS OR STERILIZE. ¿ USE PADS IMMEDIATELY AFTER OPENING. DO NOT STORE PADS IN OPENED POUCH. ¿ DO NOT ALLOW CIRCULATING WATER TO CONTAMINATE THE STERILE FIELD WHEN LINES ARE DISCONNECTED. ¿ THE ARCTIC GEL PADS SHOULD NOT BE PUNCTURED WITH SHARP OBJECTS. PUNCTURES WILL RESULT IN AIR ENTERING THE FLUID PATHWAY AND MAY REDUCE PERFORMANCE. ¿ IF WARRANTED, USE PRESSURE RELIEVING OR PRESSURE REDUCING DEVICES UNDER THE PATIENT TO PROTECT FROM SKIN INJURY. ¿ THE ARCTIC GEL PADS ARE ONLY FOR USE WITH AN ARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM. ¿ THE WATER CONTENT OF THE HYDROGEL AFFECTS THE PADS ADHESION TO THE SKIN AND CONDUCTIVITY, AND THEREFORE, THE EFFICIENCY OF CONTROLLING PATIENT TEMPERATURE. PERIODICALLY CHECK THAT PADS REMAIN MOIST AND ADHERENT. REPLACE PADS WHEN THE HYDROGEL NO LONGER UNIFORMLY ADHERES TO THE SKIN. REPLACING PADS AT LEAST EVERY 5 DAYS IS RECOMMENDED. ¿ IF NEEDED, PLACE DEFIBRILLATION PADS BETWEEN THE ARCTIC GEL PADS AND THE PATIENT¿S SKIN. ¿ DISCARD USED ARCTIC GEL PADS IN ACCORDANCE WITH HOSPITAL PROCEDURES FOR MEDICAL WASTE. DIRECTIONS FOR USE: 1. ARCTIC GEL PADS ARE ONLY FOR USE WITH AN ARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM CONTROL MODULE. SEE OPERATORS MANUAL FOR DETAILED INSTRUCTIONS ON SYSTEM USE. 2. SELECT THE PROPER NUMBER, SIZE AND STYLE PAD FOR THE PATIENT SIZE AND CLINICAL INDICATION. HOWEVER, THE RATE OF TEMPERATURE CHANGE AND POTENTIALLY THE FINAL ACHIEVABLE TEMPERATURE IS AFFECTED BY PAD SURFACE AREA, PATIENT SIZE, PAD PLACEMENT AND WATER TEMPERATURE RANGE. BEST SYSTEM PERFORMANCE WILL BE ACHIEVED BY USING THE ENTIRE PAD SET. IF THE ENTIRE SET OF PADS IS NOT USED, THE MINIMUM FLOW RATE MAY NOT BE ACHIEVED. 3. FOR PATIENT COMFORT, THE PADS MAY BE PREWARMED USING WATER TEMPERATURE CONTROL MODE (MANUAL) PRIOR TO APPLICATION. 4. PLACE THE PADS ON HEALTHY, CLEAN SKIN ONLY. REMOVE ANY CREAMS OR LOTIONS FROM PATIENT¿S SKIN BEFORE PAD APPLICATION. REMOVE THE RELEASE LINER FROM EACH PAD AND APPLY TO THE APPROPRIATE AREA. THE PADS MAY BE OVERLAPPED OR FOLDED ADHESIVE TO ADHESIVE TO ACHIEVE PROPER PLACEMENT. THE PADS MAY BE REMOVED AND REAPPLIED IF NECESSARY. THE PAD SURFACE MUST BE CONTACTING THE SKIN FOR OPTIMAL ENERGY TRANSFER EFFICIENCY. PLACE PADS TO ALLOW FOR FULL RESPIRATORY EXCURSION. 5. ATTACH THE PAD¿S LINE CONNECTORS TO THE PATIENT LINE MANIFOLDS. BEGIN CIRCULATING WATER THROUGH THE PADS USING EITHER PATIENT TEMPERATURE CONTROL MODE (AUTOMATIC) OR WATER TEMPERATURE CONTROL MODE (MANUAL). IF THE PADS FAIL TO PRIME OR A SIGNIFICANT CONTINUOUS AIR LEAK IS OBSERVED IN THE PAD RETURN LINE, CHECK CONNECTIONS, THEN IF NEEDED REPLACE THE LEAKING PAD. 6. ONCE THE PADS ARE PRIMED, ASSURE THE FLOW RATE DISPLAYED ON THE CONTROL PANEL IS GREATER THAN 1.7 LITERS PER MINUTE, WHICH IS THE MINIMUM FLOW RATE FOR A FULL PAD KIT . 7. WHEN FINISHED, EMPTY WATER FROM PADS. COLD TEMPERATURE INCREASES THE ADHESIVENESS OF THE HYDROGEL. FOR EASE OF REMOVAL, LEAVE PADS ON THE PATIENT FOR APPROXIMATELY 15 MINUTES TO ALLOW THE HYDROGEL TO WARM. SLOWLY REMOVE PADS FROM THE PATIENT AND DISCARD." CORRECTION: D H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. H3 OTHER TEXT : THE ACTUAL/SUSPECTED DEVICE WAS INSPECTED
IT WAS REPORTED THAT THE WATER FLOW THROUGH THE ARCTIC GEL PAD WAS STOPPED AFTER THE ONSET OF HYPOTHERMIA. THE LOW FLOW CHIMNEY FLASHED ON THE DISPLAY. BY ELIMINATING THE INDIVIDUAL CAUSES IT TURNED OUT THAT ONE LINE SUPPLYING THE WATER TO THE PAD WAS BLOCKED. PER ADDITIONAL INFORMATION RECEIVED FROM THE IBC VIA EMAIL ON (B)(6) 2021 THE PROCEDURE WAS COMPLETED WITH THE USE OF ANOTHER ARCTIC GEL NEONATAL PAD.
THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT THE FLOW OF WATER THROUGH THE ARCTIC GEL PAD STOPPED AFTER THE ONSET OF HYPOTHERMIA. THE LOW FLOW CHIMNEY FLASHED ON THE DISPLAY. ELIMINATING THE INDIVIDUAL CAUSES, IT TURNED OUT THAT ONE LINE SUPPLYING THE WATER TO THE PAD WAS BLOCKED. PER ADDITIONAL INFORMATION RECEIVED VIA IBC ON 08JAN2021, THE PROCEDURE WAS COMPLETED WITH USE OF OTHER ARCTIC GEL NEONATE PAD.