SYRINGE 60ML E/T
Report
- Report Number
- 3003152976-2021-00043
- Event Type
- Malfunction
- Date Received
- January 20, 2021
- Date of Event
- December 28, 2020
- Report Date
- January 29, 2021
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- UDI-DI
- 30382903008668
- PMA / PMN Number
- K110771
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES. D.10. RETURNED TO MANUFACTURER ON: 1/22/2021. H.6. INVESTIGATION: ONE SEALED SAMPLE OF LOT 1911289 WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THE PRODUCT WAS VISUALLY INSPECTED AND THE TIP IS OBSERVED TO BE BENT. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2003201, FINDING ONE ANNOTATION RELATED TO THE ALLEGED DEFECT. DURING THE MARKING PROCESS, A FAILURE WAS DETECTED RELATED TO AN INCREASED SPEED IN THE CONVEYOR BELT THAT TRANSFERS THE SYRINGES TO THE MARKING MACHINE WHICH RESULTED IN PRODUCT GETTING DAMAGED. THE MECHANICAL TEAM REPAIRED THE FAILURE AND IMPACTED UNITS WERE SCRAPPED. IT WAS DETERMINED THE ISSUE REPORTED IS RELATED TO THIS MALFUNCTION.
IT WAS REPORTED THAT A SYRINGE 60ML E/T HAD THE TIP DAMAGED AT AN UNSPECIFIED TIME. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "THE TIP BENT".
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT A SYRINGE 60ML E/T HAD THE TIP DAMAGED AT AN UNSPECIFIED TIME. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "THE TIP BENT".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 97909 | SYRINGE 60ML E/T | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 300866 | 1911289 | 30382903008668 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |