FDA Adverse Event Malfunction Summary report: N

FEATHERLITE FLOOD SOURCE

MDR report key: 1120112 · Received August 15, 2008

Report

Report Number
2029030-2008-00001
Event Type
Malfunction
Date Received
August 15, 2008
Date of Event
July 16, 2008
Report Date
August 15, 2008
Manufacturer
NORTH AMERICAN SCIENTIFIC
Product Code
IXD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

WHILE THE MANUFACTURER BELIEVES THE LIKELIHOOD OF SERIOUS INJURY FROM A RECURRENCE OF THIS EVENT IS REMOTE, THE FOLLOWING ACTIONS HAVE BEEN TAKEN. THE VENDOR'S INSPECTION OF THE RETURNED SHIELDING CASE SHOWED THAT THE BOND ON ONE OF THE HINGE ATTACHMENTS WAS INADEQUATE. IT IS BELIEVED THAT WHEN THE FIRST HINGE JOINT FAILED, EXCESSIVE FORCE WAS PLACED ON THE REMAINING HINGE JOINTS RESULTING IN THE FAILURE. THE VENDOR HAS IMPLEMENTED A SECONDARY BONDING STEP DURING FINAL MANUFACTURING PROCESS USING A LOW VISCOSITY WICKING ADHESIVE TO ENSURE THAT ALL HINGE POINTS ARE PROPERLY BONDED.

Description of Event or Problem · 1

THE HANDLE FOR A FLOOD SOURCE SHIELDED CASE BECAME DETACHED WHILE THE TECHNICIAN WAS TRANSFERRING THE SOURCE AND CASE TO THE (B) (4) CAMERA. THE CASE FELL ON HER CALF. THE INJURED TECHNICIAN WALKED TO THE ER FOR TREATMENT. SHE WAS GIVEN PAIN MEDICATION; HER LEG WAS WRAPPED IN AN ACE BANDAGE AND ICE, AND WAS TOLD TO STAY OFF HER FEET FOR LONG PERIODS. THE ER REPORT WARNS THAT MUSCLE CONTUSIONS CAUSE BLEEDING IN THE DEEP TISSUES THAT CAN INCREASE THE PRESSURE IN THE EXTREMITY. IT ALSO STATES THAT THERE IS A SMALL CHANCE THAT THE PRESSURE FROM THE BLEEDING AND SWELLING WILL INTERFERE WITH THE CIRCULATION. IF THIS HAPPENS, PERMANENT DAMAGE TO THE MUSCLE CAN OCCUR WITHIN 4-8 HOURS IF UNTREATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FEATHERLITE FLOOD SOURCE NORTH AMERICAN SCIENTIFIC IXD NORTH AMERICAN SCIENTIFIC MED3709

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention