LIFEPAK 20 DEFIBRILLATOR/MONITOR

MDR report key: 1120069 · Received August 15, 2008

Report

Report Number
3015876-2008-00949
Event Type
Malfunction
Date Received
August 15, 2008
Date of Event
July 16, 2008
Report Date
July 16, 2008
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
LDD
PMA / PMN Number
K012274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. IT WAS CONFIRMED THAT THE SYSTEM CONTROLLER PCB WAS CAUSING THE DEVICE TO CONTINUALLY RESET. PHYSIO-CONTROL REPLACED THE SYSTEM CONTROLLER PCB, AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD REBOOT ON ITS OWN. THERE WERE NO REPORTS OF PT USE ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 20 DEFIBRILLATOR/MONITOR LDD PHYSIO-CONTROL, INC. 20 NA

Patients

Seq Age Sex Outcome Treatment
1 NA