FDA Adverse Event Malfunction Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1120023 · Received August 14, 2008

Report

Report Number
2023826-2008-01070
Event Type
Malfunction
Date Received
August 14, 2008
Date of Event
June 5, 2008
Report Date
July 28, 2008
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND THE LENS OPTIC TORN. PIECES OF THE LENS OPTIC AND ONE HAPTIC WERE TORN OFF. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. IT SHOULD BE NOTED THAT THE INJECTOR AND CARTRIDGE WERE NOT RETURNED FOR EVALUATION. CONCLUSIONS - A MULTIFUNCTIONAL TEAM INVESTIGATED COMPLAINTS REGARDING LENS TEARS ASSOCIATED WITH THE CQ CARTRIDGE: THE RESULTANT CORRECTIVE ACTIONS INCLUDED IMPROVEMENTS IN MANUFACTURING, RELEASE TESTING, AND EVALUATION OF TEST RESULTS. ADDITIONALLY, THE INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

THE REPORTER STATED THE SURGEON INSERTED A CQ2015A COLLAMER ASPHERIC THREE PIECE LENS AND THE HAPTIC TORE. THE LENS WAS REMOVED WITH NO REPORT OF ANY PATIENT INJURY. ANOTHER SAME TYPE LENS WAS INSERTED BUT THIS LENS ALSO TORE - SEE MFR REPORT # 2023826-2008-01071. A THIRD DIFFERENT TYPE LENS WAS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY CQ2015A NA

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE MOEDL: CQ CARTRIDGE-FP| INJECTOR MODEL: MSI-PM