MICROSTAAR FOAM TIP PLUNGER
Report
- Report Number
- 2023826-2008-01061
- Event Type
- Malfunction
- Date Received
- August 14, 2008
- Date of Event
- July 22, 2008
- Report Date
- July 23, 2008
- Manufacturer
- STAAR SURGICAL COMPANY
- Product Code
- MSS
- PMA / PMN Number
- K980696
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- NURSE
Narratives
THE PRODUCT PERTAINING TO THIS COMPLAINT WAS NOT RETURNED, HOWEVER, THE LENS WAS RETURNED AND EVALUATED. A HAPTIC WAS TORN OFF AND THERE WAS CLEAR SURGICAL RESIDUE ON THE LENS. CONCLUSIONS: AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN 2005. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CANNOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.
IT WAS REPORTED THAT THE SURGEON INSERTED A CC4204BF COLLAMER PLATE LENS AND THE FOAM TIP PLUNGER DID NOT SPREAD OPEN AND TORE THE LENS. THE LENS WAS REMOVED WITHOUT ANY PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICROSTAAR FOAM TIP PLUNGER | INTRAOCULAR LENS FOLDERS AND INJECTORS | MSS | STAAR SURGICAL COMPANY | FOAM TIP PLUNGER | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | LENS MODEL: CC4204BF| INJECTOR MODEL: UNKNOWN| CARTRIDGE MODEL: SFC-25 FP |