FDA Adverse Event Malfunction Summary report: N

CATALYS SYSTEM

MDR report key: 11197599 · Received January 19, 2021

Report

Report Number
3006695864-2021-07050
Event Type
Malfunction
Date Received
January 19, 2021
Date of Event
December 28, 2020
Report Date
March 1, 2021
Manufacturer
AMO MANUFACTURING USA, LLC
Product Code
OOE
UDI-DI
05050474609792
PMA / PMN Number
K200056
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

G4. PMA/510(K) NUMBER CORRECTED TO: K200056. H3. DEVICE EVALUATED BY MANUFACTURER: UPDATED. H4. DEVICE MANUFACTURE DATE CORRECTED TO: JUN 29, 2016. H6. CLINICAL CODE CORRECTED TO 4582 AND FIELDS UPDATED. A RECORD REVIEW WAS PERFORMED. A PRODUCT DEFICIENCY REVIEW WAS PERFORMED AND THERE IS NO PRODUCT DEFICIENCY IDENTIFIED. A DOCUMENT, SERVICE HISTORY, AND TRENDING WAS REVIEWED. THERE IS NOT A RECOGNIZABLE ADVERSE TREND. THE RISKS AND MITIGATIONS ASSOCIATED WITH THE COMPLAINT ISSUE ARE IDENTIFIED IN EXISTING RISK DOCUMENTS AND NO NEW RISKS WERE IDENTIFIED AS PART OF THIS INVESTIGATION. A LABELING REVIEW WAS CONDUCTED; THE OPERATOR MANUAL FOR THE SYSTEM WAS REVIEWED AND FOUND TO INCLUDE ADEQUATE INSTRUCTIONS FOR USE, WARNINGS AND OPERATIONAL ERRORS. THE REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR CATALYS-U LASER SYSTEM S/N (B)(6) SHOWED THAT THERE WERE NO ISSUES OR NON-CONFORMITIES. THE SYSTEM AND ITS COMPONENTS MET ALL SPECIFICATIONS PRIOR TO BEING RELEASED. MANUFACTURING HAS BEEN RULED OUT AS A POTENTIAL CAUSE FOR THE REPORTED ISSUE. BASED ON THE INVESTIGATION RESULTS, NO CORRECTIVE ACTION HAS BEEN ISSUED. BASED ON THE INVESTIGATION RESULTS THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. JOHNSON & JOHNSON SURGICAL VISION WILL CONTINUE TO MONITOR THIS TYPE OF COMPLAINTS ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 1

A REVIEW OF RECORDS RELATED TO THE DEVICE INCLUDING LABELING, COMPLAINT TRENDING, AND RISK DOCUMENTATION WILL BE PERFORMED. UPON COMPLETION OF THIS REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 1

CUSTOMER REPORTED SUCTION LOSS DURING LASER FIRING. THERE WAS FLUID LOSS DURING TREATMENT, BUT NO PATIENT MOVEMENT. NO REPORTED PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
89183 CATALYS SYSTEM OPHTHALMIC FEMTOSECOND LASER OOE AMO MANUFACTURING USA, LLC CATALYS-U 05050474609792

Patients

Seq Age Sex Outcome Treatment
1 LIQUID OPTICS INTERFACE, SIZE 14, GEN2