DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2021-12913
- Event Type
- Malfunction
- Date Received
- January 19, 2021
- Date of Event
- August 22, 2020
- Report Date
- January 20, 2021
- Manufacturer
- DEXCOM, INC.
- Product Code
- QBJ
- UDI-DI
- 00386270000385
- PMA / PMN Number
- DEN170088
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
B5 --DATE RECEIVED BY MANUFACTURER CORRECTED IN G3 TO 01/17/2021 AS THE WRONG YEAR WAS RECORDED. G3 DATE RECEIVED BY MANUFACTURER CORRECTED TO 01/17/2021.
DATE RECEIVED BY MANUFACTURER CORRECTED IN G3 TO 01/17/2021 AS THE WRONG YEAR WAS RECORDED.
(B)(4).
IT WAS REPORTED THAT A PAIRING FAILURE OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. AN EXTERIOR PHYSICAL VISUAL INSPECTION WAS PERFORMED AND PASSED. A TRANSMITTER VOLTAGE TEST WAS PERFORMED AND FAILED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND A PAIRING FAILURE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. IN ADDITION, TRANSMITTER FAILED ERROR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE REPORTED EVENT OF A PAIRING FAILURE IS REPORTABLE BASED ON THE FINDING OF A TRANSMITTER FAILED ERROR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 92305 | DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | QBJ | DEXCOM, INC. | 9445-02 | 7271269 | 00386270000385 |