FDA Adverse Event Injury Summary report: N

DELTECH

MDR report key: 111821 · Received July 10, 1997

Report

Report Number
111821
Event Type
Injury
Date Received
July 10, 1997
Date of Event
September 8, 1996
Report Date
April 11, 1997
Manufacturer
DELTECH ENGINEERING CENTURY PARK
Product Code
FRF
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
MT, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE DRYER ON THE MEDICAL AIR SUPPLY SYSTEM FAILED TO REMOVE WATER AND THE LINE MONITOR FAILED TO ALARM. THE WATER IN THE SYSTEM CAUSED THE FLOW SENSOR ON THE VENTILATOR TO ALARM. HOWEVER THE VENTILATOR CONTINUED TO OPERATE. THE VENTILATOR WAS EXCHANGED WITHOUT PT COMPROMISE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DELTECH MEDICAL AIR DRYER FRF DELTECH ENGINEERING CENTURY PARK 8M50 NA

Patients

Seq Age Sex Outcome Treatment
1 53 YR Required Intervention NELTCOR, PURITAN BENNETT 7200 AE VENTILATOR| SN 9043077