FDA Adverse Event
Injury
Summary report: N
DELTECH
MDR report key: 111821
·
Received July 10, 1997
Report
- Report Number
- 111821
- Event Type
- Injury
- Date Received
- July 10, 1997
- Date of Event
- September 8, 1996
- Report Date
- April 11, 1997
- Manufacturer
- DELTECH ENGINEERING CENTURY PARK
- Product Code
- FRF
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- MT, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE DRYER ON THE MEDICAL AIR SUPPLY SYSTEM FAILED TO REMOVE WATER AND THE LINE MONITOR FAILED TO ALARM. THE WATER IN THE SYSTEM CAUSED THE FLOW SENSOR ON THE VENTILATOR TO ALARM. HOWEVER THE VENTILATOR CONTINUED TO OPERATE. THE VENTILATOR WAS EXCHANGED WITHOUT PT COMPROMISE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DELTECH | MEDICAL AIR DRYER | FRF | DELTECH ENGINEERING CENTURY PARK | 8M50 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Required Intervention | NELTCOR, PURITAN BENNETT 7200 AE VENTILATOR| SN 9043077 |