ALINITY C PROCESSING MODULE
Report
- Report Number
- 3002809144-2021-00031
- Event Type
- Malfunction
- Date Received
- January 15, 2021
- Date of Event
- December 20, 2020
- Report Date
- February 11, 2021
- Manufacturer
- ABBOTT GMBH
- Product Code
- JJE
- UDI-DI
- 00380740137380
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DURING THE REQUESTED SITE VISIT, THE FIELD SERVICE REPRESENTATIVE PERFORMED MULTIPLE TROUBLESHOOTING SERVICES. INCLUDING THE CLEANING OF THE CUVETTE WASHER NOZZLES, NOZZLE C4 #1, #4, #5 (ROHS) (LN 7-205228-02). WHICH RESOLVED THE ISSUE. RETURN TESTING WAS NOT COMPLETED, AS RETURNS WERE NOT AVAILABLE. A REVIEW OF THE ALINITY C PROCESSING MODULE, SERIAL NUMBER (B)(6) SERVICE HISTORY, REVEALED NO CONTRIBUTING FACTORS ON OR AROUND THE TIME OF THE ISSUE. THE ALINITY CI-SERIES OPERATIONS MANUAL PROVIDES ADEQUATE INFORMATION REGARDING THE TROUBLESHOOTING OF ERRATIC/DISCREPANT RESULTS. THE ALINITY C SERVICE DOCUMENTATION CONTAINS ADEQUATE INFORMATION FOR THE TROUBLESHOOTING OF THE NOZZLE C4 #1, #4, #5 (ROHS). A REVIEW FOR TRENDS FOR THE NOZZLE C4 #1, #4, #5 (ROHS) (LN 7-205228-02) DID NOT IDENTIFY ANY TRENDS FOR THE REPLACED PART. A PRODUCT MONITORING REVIEW OF THE ALINITY C TRACKING AND TRENDING DATA, REVEALED NO TREND FOR THE ALINITY C PLATFORM AND NO SIMILAR ISSUES FOR DISCREPANT RESULTS AS DESCRIBED IN THIS COMPLAINT. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY C PROCESSING MODULE, SN (B)(6), OR THE NOZZLE C4 #1, #4, #5 (ROHS) (LN 7-205228-02) WAS IDENTIFIED.
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THERE WAS NO ADDITIONAL PATIENT INFORMATION PROVIDED BY THE CUSTOMER DUE TO PRIVACY ISSUES.
THE CUSTOMER REPORTED FALSELY DEPRESSED SODIUM (NA) RESULTS ON FIVE PATIENTS GENERATED ON THE ALINITY C PROCESSING MODULE. THE RESULTS PROVIDED WERE FOR ONLY ONE PATIENT: NA = 121MMOL/L (NORMAL RANGE 136-145MMOL/L), 122, 101, 147, 127, 168, 140, 162, 137MMOL/L. THE CUSTOMER ALSO REPORTED ERRATIC (DEPRESSED/ELEVATED) CREATININE RESULTS ON FIVE PATIENTS GENERATED ON THE ALINITY C PROCESSING MODULE. THE RESULTS PROVIDED WERE FOR ONLY ONE PATIENT: CREATININE = 3.34MG/DL (RANGE 0.57-1.25MG/DL) / 0.89 AND 2.75MG/DL. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 75069 | ALINITY C PROCESSING MODULE | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | ABBOTT GMBH | 03R6701 | 00380740137380 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |