8100 ALARIS PUMP MODULE
Report
- Report Number
- 2016493-2021-09416
- Event Type
- Malfunction
- Date Received
- January 15, 2021
- Report Date
- April 9, 2020
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403222054
- PMA / PMN Number
- K133532
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THE CUSTOMER REPORTED PROBLEM WAS CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACK WISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
BD QUALITY ADVOCATE, THIS NOTIFICATION IS TO INFORM YOU THAT A NEW CASE HAS BEEN CREATED WITH THE COMPLAINT TYPE CATEGORY INFUSION CA. (B)(4). CASE DESCRIPTION: CALLER HAS AN 8100 MODULE GIVING HIM A 242.4030 ERROR. SN: (B)(4). FAILURE DEVICE TYPE: FAILURE PROBLEM TYPE: FAILURE MODE: CASE RESOLUTION: USED KA 12066 ALARIS INFUSION ERROR 242.4030. NO MOTOR MOVEMENT. RECOMMENDED TO REPLACE MOTOR CONTROL BOARD. REF: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 76123 | 8100 ALARIS PUMP MODULE | PUMP, INFUSION | FRN | CAREFUSION SD | 8100 | 10885403222054 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |