FDA Adverse Event Malfunction Summary report: N

PROG VALVE RIGHT ANGLE

MDR report key: 11165145 · Received January 13, 2021

Report

Report Number
3013886523-2021-00017
Event Type
Malfunction
Date Received
January 13, 2021
Date of Event
October 20, 2020
Report Date
March 16, 2021
Manufacturer
INTEGRA LIFESCIENCES SWITZERLAND SAR
Product Code
JXG
PMA / PMN Number
K053107
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE HAKIM VALVE WAS RETURNED FOR EVALUATION: DEVICE HISTORY RECORD (DHR) - LOT 3841197, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK FAILURE ANALYSIS - THE VALVE WAS VISUALLY INSPECTED, SOME BIOLOGICAL DEBRIS WAS NOTED. THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS 100MMH2O. THE VALVE WAS HYDRATED. THE VALVE PASSED THE TEST FOR PROGRAMMING, LEAKS, OCCLUSION, REFLUX AND PRESSURE. NO ROOT CAUSE COULD BE DETERMINED AS THE TECHNICIAN COULD NOT CONFIRM ANY PROBLEM WITH THE VALVE AT THE TIME OF INVESTIGATION. THE POSSIBLE ROOT CAUSE FOR THE ISSUE REPORTED BY THE CUSTOMER COULD BE BIOLOGICAL DEBRIS AND PROTEIN BUILD UP INTERFERING WITH THE VALVE MECHANISM, AT THE TIME OF THE INVESTIGATION NO OCCLUSION ISSUES WERE NOTED.

Additional Manufacturer Narrative · 1

AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

A FACILITY REPORTED A HAKIM VALVE WAS FOUND OBSTRUCTED DURING IMPLANTATION PROCEDURE. IT WAS REPLACED WITH A NEW VALVE.