IMPACTOR HEAD
Report
- Report Number
- 0008031020-2021-00021
- Event Type
- Malfunction
- Date Received
- January 13, 2021
- Date of Event
- December 21, 2020
- Report Date
- March 10, 2021
- Manufacturer
- STRYKER GMBH
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE REPORTED EVENT COULD BE CONFIRMED, SINCE THE DEVICE WAS RETURNED FOR EVALUATION AND MATCHES THE ALLEGED FAILURE MODE. DEVICE INSPECTION REVEALED THE FOLLOWING: THE IMPACTOR HEAD WAS RECEIVED BROKEN IN TWO PARTS. THE BREAKAGE, WAS JUST AROUND THE MIDDLE. THE OVERALL BREAKAGE LOOKED TO BE CLEAN. INDICATING, TOWARDS A SUDDEN BREAKAGE ALONG THE WEAKER REGIONS. HOWEVER, SURFACE AT THE POINT OF DETACHMENT SHOWED, SLIGHT MATERIAL DEFORMATION AND WAVES. SUGGESTING TOWARDS HEAVY EXTERNAL IMPACTS. A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE, ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. NO INDICATIONS OF MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE FOUND, DURING THE INVESTIGATION. A REVIEW OF THE LABELING DID NOT INDICATE, ANY ABNORMALITIES. BASED ON THE ABOVE INVESTIGATION, THE ROOT CAUSE OF THE FAILURE IS USER RELATED. THERE ARE SIGNS OF HEAVY IMPACTIONS ON THE IMPACTOR AS EVIDENT BY THE MANNER OF DEFORMATION AND BREAKAGE SURFACE. THE DEVICE IS MEANT FOR ONLY GENTLE HAMMERING. THE OP-TECH CLEARLY WARNS THE USER THAT: ¿NOTE: USE GENTLE HAMMERING ONLY. OTHERWISE THE IMPACTOR MAY BE DESTROYED¿. IF ANY FURTHER INFORMATION IS PROVIDED, THE COMPLAINT REPORT WILL BE UPDATED.
AS REPORTED: "THE 704001-1 (IMPACTOR HEAD OMEGA) BROKE, DURING AN OMEGA PLATE IMPLANTATION".
UPON COMPLETION OF INVESTIGATION, ADDITIONAL INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
AS REPORTED: "THE 704001-1 (IMPACTOR HEAD OMEGA) BROKE DURING AN OMEGA PLATE IMPLANTATION".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 57473 | IMPACTOR HEAD | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | STRYKER GMBH | 02744V |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |