FDA Adverse Event Injury Summary report: N

SYRINGE 20ML LL S/C 50

MDR report key: 11159521 · Received January 13, 2021

Report

Report Number
9614033-2021-00005
Event Type
Injury
Date Received
January 13, 2021
Date of Event
December 4, 2020
Report Date
February 8, 2021
Manufacturer
BECTON DICKINSON DE MEXICO
Product Code
FMF
PMA / PMN Number
K151766
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: IT WAS RECEIVED 3 PHOTOGRAPHS IN WHICH IT IS POSSIBLE TO OBSERVE IN EACH ONE OF THIS A SYRINGE WITH THE DAMAGE BARREL, THE PHOTOGRAPHS SHOWS THE BLISTER PAPER TOO IN WHICH ITS POSSIBLE TO OBSERVE THE CODE 303310 AND THE BATCHES 0136387 AND 0098741 THE SAME THAT THE CUSTOMER REPORTED. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. THE DAMAGE TO THE SYRINGE CAN BE GENERATED BY EXCESS SYRINGES ASSEMBLED IN THE HOPPER, AS THERE IS ACCUMULATION OF SYRINGE ASSEMBLED IN THE ELEVATOR, THIS GENERATED OBSTRUCTION AND DAMAGE TO THE SYRINGE ASSEMBLED. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE 20ML LL S/C 50 WAS CRACKED AND LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 303310 BATCH NO: 0136387, 0098741.   IT WAS REPORTED THAT 3 SYRINGES CRACKING DOWN THE RIGHT SIDE OF THE SYRINGE.   VERBATIM: REPORT OF 3 SYRINGES CRACKING DOWN THE RIGHT SIDE OF THE SYRINGE. SPRAYED TECH IN THE FACE W/ VENOFOR(?), HE SOUGHT MEDICAL ATTN, INITIALS JMM, WEEK OF (B)(6) SHE HAS THE SYRINGES AVAILABLE TO RETURN.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 20ML LL S/C 50 WAS CRACKED AND LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: (B)(4). BATCH NO: 0136387, 0098741.   IT WAS REPORTED THAT 3 SYRINGES CRACKING DOWN THE RIGHT SIDE OF THE SYRINGE.   VERBATIM: REPORT OF 3 SYRINGES CRACKING DOWN THE RIGHT SIDE OF THE SYRINGE. SPRAYED TECH IN THE FACE W/ VENOFOR(?), HE SOUGHT MEDICAL ATTN, INITIALS JMM, WEEK OF DEC 4TH SHE HAS THE SYRINGES AVAILABLE TO RETURN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
61176 SYRINGE 20ML LL S/C 50 PISTON SYRINGE FMF BECTON DICKINSON DE MEXICO 0098741

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention