SYRINGE 20ML LL S/C 50
Report
- Report Number
- 9614033-2021-00005
- Event Type
- Injury
- Date Received
- January 13, 2021
- Date of Event
- December 4, 2020
- Report Date
- February 8, 2021
- Manufacturer
- BECTON DICKINSON DE MEXICO
- Product Code
- FMF
- PMA / PMN Number
- K151766
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: IT WAS RECEIVED 3 PHOTOGRAPHS IN WHICH IT IS POSSIBLE TO OBSERVE IN EACH ONE OF THIS A SYRINGE WITH THE DAMAGE BARREL, THE PHOTOGRAPHS SHOWS THE BLISTER PAPER TOO IN WHICH ITS POSSIBLE TO OBSERVE THE CODE 303310 AND THE BATCHES 0136387 AND 0098741 THE SAME THAT THE CUSTOMER REPORTED. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. THE DAMAGE TO THE SYRINGE CAN BE GENERATED BY EXCESS SYRINGES ASSEMBLED IN THE HOPPER, AS THERE IS ACCUMULATION OF SYRINGE ASSEMBLED IN THE ELEVATOR, THIS GENERATED OBSTRUCTION AND DAMAGE TO THE SYRINGE ASSEMBLED. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT SYRINGE 20ML LL S/C 50 WAS CRACKED AND LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 303310 BATCH NO: 0136387, 0098741. IT WAS REPORTED THAT 3 SYRINGES CRACKING DOWN THE RIGHT SIDE OF THE SYRINGE. VERBATIM: REPORT OF 3 SYRINGES CRACKING DOWN THE RIGHT SIDE OF THE SYRINGE. SPRAYED TECH IN THE FACE W/ VENOFOR(?), HE SOUGHT MEDICAL ATTN, INITIALS JMM, WEEK OF (B)(6) SHE HAS THE SYRINGES AVAILABLE TO RETURN.
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT SYRINGE 20ML LL S/C 50 WAS CRACKED AND LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: (B)(4). BATCH NO: 0136387, 0098741. IT WAS REPORTED THAT 3 SYRINGES CRACKING DOWN THE RIGHT SIDE OF THE SYRINGE. VERBATIM: REPORT OF 3 SYRINGES CRACKING DOWN THE RIGHT SIDE OF THE SYRINGE. SPRAYED TECH IN THE FACE W/ VENOFOR(?), HE SOUGHT MEDICAL ATTN, INITIALS JMM, WEEK OF DEC 4TH SHE HAS THE SYRINGES AVAILABLE TO RETURN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 61176 | SYRINGE 20ML LL S/C 50 | PISTON SYRINGE | FMF | BECTON DICKINSON DE MEXICO | 0098741 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |