FDA Adverse Event Malfunction Summary report: N

AVANOS CURVED SHARP RF CANNULA, RADIOPAQUE MARKER

MDR report key: 11154981 · Received January 12, 2021

Report

Report Number
8030647-2021-00001
Event Type
Malfunction
Date Received
January 12, 2021
Report Date
March 26, 2021
Manufacturer
AVANOS MEDICAL INC.
Product Code
GXI
UDI-DI
00193493986347
PMA / PMN Number
K972846
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HISTORY RECORD FOR LOT 20042519 WAS REVIEWED AND THE PRODUCT WAS PRODUCED ACCORDING TO PRODUCT SPECIFICATIONS. THE ACTUAL COMPLAINT PRODUCT WAS NOT RETURNED FOR EVALUATION. ROOT CAUSE COULD NOT BE DETERMINED. ALL INFORMATION REASONABLY KNOWN AS OF 24 MAR 2021 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT/REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC.. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL INC. COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD IS IN-PROGRESS. THE SAMPLE IS REPORTED TO BE AVAILABLE, BUT HAS NOT YET BEEN RECEIVED BY THE MANUFACTURER. ALL INFORMATION REASONABLY KNOWN AS OF (B)(6) 2021 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC.. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL INC. COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RADIOPAQUE PORTION OF THE CANNULA BROKE OFF DURING THE PROCEDURE AND EMBEDDED INTO THE SOFT TISSUE OF THE CERVICAL REGION. ADDITIONAL INFORMATION RECEIVED 29-DEC-2020 FROM THE PHYSICIAN INDICATED "RADIOPAQUE INDICATOR AT THE BASE OF THE ACTIVE TIP WAS NOTED TO BE DISENGAGED FROM THE CANNULA DURING PLACEMENT/MANIPULATION OF THE TIP TO GET IT IN THE RIGHT POSITION. THE STYLET WAS ENGAGED DURING MANIPULATION. RADIOPAQUE DOT COULD BE SEEN ON X-RAY DURING THE PROCEDURE THAT WAS LEFT BEHIND ALONG THE LATERAL CERVICAL WAIST." COMPONENT REPORTED TO STILL BE IN PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOT YET RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
52406 AVANOS CURVED SHARP RF CANNULA, RADIOPAQUE MARKER STANDARD RF PROBES AND CANNULAS GXI AVANOS MEDICAL INC. PMF20-100-10CS 20042519 00193493986347

Patients

Seq Age Sex Outcome Treatment
1 52 YR