FDA Adverse Event Injury Summary report: N

REVERSE SHOULDER SYSTEM GLENOSPHERE 39X24.5

MDR report key: 11153954 · Received January 12, 2021

Report

Report Number
3005180920-2021-00045
Event Type
Injury
Date Received
January 12, 2021
Date of Event
December 24, 2020
Report Date
February 26, 2021
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
PHX
UDI-DI
07630040706575
PMA / PMN Number
K170452
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE VISUAL INSPECTION WAS PERFORMED ON WEDNESDAY, 27TH JANUARY 2021 BY MEDACTA R&D DEPARTMENT. THE LINER DOES NOT PRESENT ANY SIGNS OF DAMAGE. THE GLENOSPHERE IS PARTIALLY SCRATCHED AND DULL, PROBABLY DUE TO FRICTION WITH THE "REVERSE METAPHYSIS" OCCURRED AFTER THE DISLOCATION. THE REVERSE METAPHYSIS SHOWS MINOR BONE AND HYDROXYAPATITE RESIDUALS ON THE COATED SURFACE. THE OUTER CYLINDRIAL SURFACE IS DAMAGED ON THE MEDIAL SIDE PRESUMABLY DUE TO CONTACT WITH THE GLENOSPHERE.

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 12-JAN-2021: LOT 1910953: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 04-JUN-2020. EXPIRATION DATE: 24-MAY-2025. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. ADDITIONAL IMPLANT INVOLVED: REVERSE SHOULDER SYSTEM 04.01.0124 HUMERAL REVERSE HC LINER Ø39/+6MM (K170452) LOT. 1907768. BATCH REVIEW PERFORMED ON 12 JANUARY 2021: LOT 1907768: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 07-NOV-2019. EXPIRATION DATE: 2024-10-16. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITH ANOTHER SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

SECOND REVISION SURGERY DUE TO SHOULDER LUXATION (INSTABILITY) OF THE GLENOSPHERE FROM THE LINER, 1 MONTH AFTER THE PREVIOUS OPERATION. DURING THE REVISION SURGERY THE METAPHYSIS, GLENOSPHERE AND LINER WERE REVISED. THE NEW GLENOSPHERE IS LATERALIZED AND THE LINER IS +3MM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
54093 REVERSE SHOULDER SYSTEM GLENOSPHERE 39X24.5 REVERSE SHOULDER GLENOSPHERE PHX MEDACTA INTERNATIONAL SA 04.01.0170 1910953 07630040706575

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention