SKYTRON
Report
- Report Number
- 1825014-2021-00003
- Event Type
- Malfunction
- Date Received
- January 11, 2021
- Date of Event
- October 22, 2020
- Report Date
- January 11, 2021
- Manufacturer
- GCX CORPORATION
- Product Code
- FQO
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- NJ, US
- Reporter Occupation
- 003
Narratives
THE VST ACCESSORY CLAMP WAS STRIPPED WHERE THE GCX VHM MONITOR ARM ATTACHES TO THE SCREWS. THIS WAS CAUSED BY AN OVER ADJUSTMENT OF THE 4 SET SCREWS THAT KEEP THE GCX VHM MONITOR ARM FIRMLY ATTACHED TO THE VST CLAMP. THE SET SCREWS WERE SO FIRMLY ADJUSTED THAT THEY CARVED GROOVES INTO THE STAINLESS STEEL HOUSING AND THEN CAUSED 2 OF THE SET SCREWS TO BE STRIPPED AND ARE UNABLE TO BE REMOVED AS INTENDED. THIS OVER ADJUSTMENT IN ADDITION TO TRAUMA TO THE ASSEMBLY AS A WHOLE IS WHAT CAUSED THE FAILURE TO OCCUR. I WAS ABLE TO ENSURE THAT THE MALFUNCTION DID NOT EFFECT THE MONITOR OR MONITOR WIRING. THE MONITOR STILL SHOWS PICTURE AND FUNCTIONS PROPERLY. OR 20 SN (B)(4) I REPLACED THE VST ACCESSORY CLAMP, (H830039A), ON THE VST. I ATTACHED THE GCX MOUNTING BLOCK FOR CLAMP, (H830040B) AND SECURED A NEW GCX VARIABLE HEIGHT ARM, (H801050A), WITH DOWN POST KIT, (H80105001). I ATTACHED THE MONITOR AND RERAN THE CABLES AND MADE THE APPROPRIATE CONNECTIONS. I SUCCESSFULLY TESTED THE SIGNAL TO THE MONITOR AND ENSURED IT WAS SECURE.
SKYTRON COMPLAINT NUMBER: (B)(4). A MONITOR AND MONITOR BRACKET DETACHED FROM THE VST CLAMP THAT IT MOUNTS TO. THE VST ACCESSORY CLAMP WAS STRIPPED WHERE THE GCX VHM MONITOR ARM ATTACHES TO THE SCREWS. THIS WAS CAUSED BY AN OVER ADJUSTMENT OF THE 4 SET SCREWS THAT KEEP THE GCX VHM MONITOR ARM FIRMLY ATTACHED TO THE VST CLAMP. THE SET SCREWS WERE SO FIRMLY ADJUSTED THAT THEY CARVED GROOVES INTO THE STAINLESS STEEL HOUSING AND THEN CAUSED 2 OF THE SET SCREWS TO BE STRIPPED AND ARE UNABLE TO BE REMOVED AS INTENDED. THIS OVER ADJUSTMENT IN ADDITION TO TRAUMA TO THE ASSEMBLY AS A WHOLE IS WHAT CAUSED THE FAILURE TO OCCUR. I WAS ABLE TO ENSURE THAT THE MALFUNCTION DID NOT EFFECT THE MONITOR OR MONITOR WIRING. THE MONITOR STILL SHOWS PICTURE AND FUNCTIONS PROPERLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 47476 | SKYTRON | VHM WALL MOUNT NEW 8/20.3 EXT. EXTENSION ARM | FQO | GCX CORPORATION | H8-010-50-A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |