FDA Adverse Event Malfunction Summary report: N

SKYTRON

MDR report key: 11147013 · Received January 11, 2021

Report

Report Number
1825014-2021-00003
Event Type
Malfunction
Date Received
January 11, 2021
Date of Event
October 22, 2020
Report Date
January 11, 2021
Manufacturer
GCX CORPORATION
Product Code
FQO
Product Problem
Yes
Report Source
Distributor report
Reporter Location
NJ, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE VST ACCESSORY CLAMP WAS STRIPPED WHERE THE GCX VHM MONITOR ARM ATTACHES TO THE SCREWS. THIS WAS CAUSED BY AN OVER ADJUSTMENT OF THE 4 SET SCREWS THAT KEEP THE GCX VHM MONITOR ARM FIRMLY ATTACHED TO THE VST CLAMP. THE SET SCREWS WERE SO FIRMLY ADJUSTED THAT THEY CARVED GROOVES INTO THE STAINLESS STEEL HOUSING AND THEN CAUSED 2 OF THE SET SCREWS TO BE STRIPPED AND ARE UNABLE TO BE REMOVED AS INTENDED. THIS OVER ADJUSTMENT IN ADDITION TO TRAUMA TO THE ASSEMBLY AS A WHOLE IS WHAT CAUSED THE FAILURE TO OCCUR. I WAS ABLE TO ENSURE THAT THE MALFUNCTION DID NOT EFFECT THE MONITOR OR MONITOR WIRING. THE MONITOR STILL SHOWS PICTURE AND FUNCTIONS PROPERLY. OR 20 SN (B)(4) I REPLACED THE VST ACCESSORY CLAMP, (H8­300­39­A), ON THE VST. I ATTACHED THE GCX MOUNTING BLOCK FOR CLAMP, (H8­300­40­B) AND SECURED A NEW GCX VARIABLE HEIGHT ARM, (H8­010­50­A), WITH DOWN POST KIT, (H8­010­50­01). I ATTACHED THE MONITOR AND RE­RAN THE CABLES AND MADE THE APPROPRIATE CONNECTIONS. I SUCCESSFULLY TESTED THE SIGNAL TO THE MONITOR AND ENSURED IT WAS SECURE.

Description of Event or Problem · 1

SKYTRON COMPLAINT NUMBER: (B)(4). A MONITOR AND MONITOR BRACKET DETACHED FROM THE VST CLAMP THAT IT MOUNTS TO. THE VST ACCESSORY CLAMP WAS STRIPPED WHERE THE GCX VHM MONITOR ARM ATTACHES TO THE SCREWS. THIS WAS CAUSED BY AN OVER ADJUSTMENT OF THE 4 SET SCREWS THAT KEEP THE GCX VHM MONITOR ARM FIRMLY ATTACHED TO THE VST CLAMP. THE SET SCREWS WERE SO FIRMLY ADJUSTED THAT THEY CARVED GROOVES INTO THE STAINLESS STEEL HOUSING AND THEN CAUSED 2 OF THE SET SCREWS TO BE STRIPPED AND ARE UNABLE TO BE REMOVED AS INTENDED. THIS OVER ADJUSTMENT IN ADDITION TO TRAUMA TO THE ASSEMBLY AS A WHOLE IS WHAT CAUSED THE FAILURE TO OCCUR. I WAS ABLE TO ENSURE THAT THE MALFUNCTION DID NOT EFFECT THE MONITOR OR MONITOR WIRING. THE MONITOR STILL SHOWS PICTURE AND FUNCTIONS PROPERLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
47476 SKYTRON VHM WALL MOUNT NEW 8/20.3 EXT. EXTENSION ARM FQO GCX CORPORATION H8-010-50-A

Patients

Seq Age Sex Outcome Treatment
1