FDA Adverse Event Malfunction Summary report: N

8100 ALARIS PUMP MODULE

MDR report key: 11144369 · Received January 10, 2021

Report

Report Number
2016493-2021-02892
Event Type
Malfunction
Date Received
January 10, 2021
Report Date
January 22, 2020
Manufacturer
CAREFUSION SD
Product Code
FRN
PMA / PMN Number
K133532
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER¿S REPORTED PROBLEM WAS CONFIRMED. A DHR (DEVICE HISTORY RECORD) REVIEW CANNOT BE COMPLETED AS THE SERIAL NUMBER WAS NOT OBTAINED UPON RECEIPT OF THE COMPLAINT. ADDITIONALLY, A HISTORICAL REVIEW OF COMPLAINTS IN TRACKWISE CANNOT BE CONDUCTED. THE CUSTOMER STATED THAT THERE IS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

CASE #: 00968577 CASE SUBJECT: NPI PATIENT SIDE OCCLUSION TEST FAIL ACCOUNT NAME: TENNESSEE VALLEY VETERANS AFFAIRS HEALTHCARE SYSTEM 626A4 ACCOUNT #: 3894778 ASSET NAME: 8015LS PCU DOM V9.33.1.2 A/B/G/N ASSET LOCATION: CONTACT: KEITH TODD CONTACT EMAIL: [email protected] CONTACT PHONE: 615-895-1103 CONTACT MOBILE: PATIENT OR USER INVOLVEMENT: NO PATIENT OR USER HARM: NO CASE DESCRIPTION: TECH CALLED FOR HELP ON SERVERAL 8100 UNITS FAILURE DEVICE TYPE: FAILURE PROBLEM TYPE: FAILURE MODE: CASE RESOLUTION: TECH CALLED IN THEY HAVE 139 8100 UNITS AND THEY DOING THIER YEARLY PM TEST AND 10 OF THE 8100 UNITS ALL FAIL ON THE PATIENT SIDE OCCLUSION TEST FAIL DOING PM TEST HE HAD ALSO RAN CALIBRATION TEST ALSO ON UNIT ALL FAIL PRESSURE TEST. TECH NEED TO REPLACE PRESSURE SENSOR WHICH IS THE BOTTOM SENSOR BUT THEY CAN REPLACE BOTH WANT HURT BUT UP TO YOU GUYS, ALSO ONE 8100 UNIT THE DISPLAY BOARD IS OUT AND NEED TO BE REPLACE, CONFIRM BOTH PART NUMBERS FOR PRESSURE SENSOR & DISPLAY BOARD, TRANSFER TECH TO ORDER MGR. DEPT. TO GET PRICE QUOTE WITH TAX. TECH THANK ME FOR MY ASSIST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
41076 8100 ALARIS PUMP MODULE PUMP, INFUSION FRN CAREFUSION SD 8100

Patients

Seq Age Sex Outcome Treatment
1