LENS 4K A/C CAMERA HEAD
Report
- Report Number
- 1643264-2021-00092
- Event Type
- Malfunction
- Date Received
- January 9, 2021
- Date of Event
- December 17, 2020
- Report Date
- March 8, 2021
- Manufacturer
- SMITH & NEPHEW, INC.
- Product Code
- GCJ
- UDI-DI
- 00885556720790
- PMA / PMN Number
- K191177
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H3, H6: THE REPORTED DEVICE, INTENDED FOR USE IN TREATMENT, WAS RETURNED TO THE DESIGNATED COMPLAINT UNIT FOR INDEPENDENT EVALUATION. THERE WAS A RELATIONSHIP FOUND BETWEEN THE RETURNED DEVICE AND THE REPORTED INCIDENT. VISUAL INSPECTION OF THE RETURNED DEVICE DID NOT IDENTIFY ANY ISSUES. FUNCTIONAL EVALUATION REVEALED THAT THERE WAS LOSS OF IMAGE WHEN THE CONNECTOR IS MOVED. THE COMPLAINT HAS BEEN CONFIRMED AND THE ROOT CAUSE HAS BEEN ASSOCIATED WITH AN ELECTRICAL COMPONENT FAILURE. FACTORS, UNRELATED TO THE DESIGN OR MANUFACTURE OF THE DEVICE WHICH COULD HAVE CONTRIBUTED TO THE COMPLAINT EVENT, INCLUDE A DAMAGED CONNECTOR. A REVIEW OF THE DEVICE HISTORY RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE LOT DID NOT MEET MANUFACTURING SPECIFICATION OR WOULD NOT BE ABLE TO PERFORM AS INTENDED. A COMPLAINT HISTORY REVIEW CONCLUDED THIS WAS A REPEAT ISSUE. NO CONTAINMENT OR CORRECTIVE ACTIONS ARE RECOMMENDED AT THIS TIME.
INTERNAL COMPLAINT REFERENCE CASE (B)(4).
IT WAS REPORTED THAT THE LENS 4K A/C CAMERA HEAD PRESENTED A BLACK SCREEN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 40251 | LENS 4K A/C CAMERA HEAD | LAPAROSCOPE, GENERAL & PLASTIC SURGERY | GCJ | SMITH & NEPHEW, INC. | 72205058 | 00885556720790 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |