V60 VENTILATOR
Report
- Report Number
- 2031642-2021-00097
- Event Type
- Malfunction
- Date Received
- January 8, 2021
- Report Date
- December 14, 2020
- Product Code
- MNT
- PMA / PMN Number
- K082660
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER
Narratives
G4:10MAR2021. B4:13MAR2021. THE CUSTOMER EVALUATED THE DEVICE WITH ASSISTANCE FROM A PHILIPS REMOTE SERVICE ENGINEER (RSE), AND IT WAS DETERMINED THAT THE POWER SWITCH OVERLAY NEEDED TO BE REPLACED TO RETURN THE DEVICE TO WORKING CONDITION. THE CUSTOMER¿S BIO-MED REPLACED THE POWER SWITCH OVERLAY TO RESOLVE THE ISSUE AND BRING THE DEVICE BACK TO FUNCTIONALITY. A SIMILAR INVESTIGATION WAS PERFORMED WAS DETERMINED TO BE DELAMINATION AND CRACKING OF THE ENCAPSULATE EPOXY ENCASING THE LIGHT EMITTING DIODE (LED) DICE, AND PHYSICAL SHEARING OF THE EPOXY OR TRACE CAUSING OPENED AND INTERMITTED TRACE CONTINUITY TO THE LED. SUBMISSION OF A REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, DISTRIBUTOR, MANUFACTURER, OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
DATE OF EVENT: (B)(6) 2020. DATE OF REPORT: 08JAN2021.
IT WAS REPORTED TO PHILIPS THAT THE DEVICE HAD A LED (LIGHT EMITTING DIODE) FAILURE. THE DEVICE WAS NOT BEING USED ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO PATIENT OR USER HARM REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 33979 | V60 VENTILATOR | VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE | MNT | V60 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |