FDA Adverse Event Malfunction Summary report: N

V60 VENTILATOR

MDR report key: 11142976 · Received January 8, 2021

Report

Report Number
2031642-2021-00097
Event Type
Malfunction
Date Received
January 8, 2021
Report Date
December 14, 2020
Product Code
MNT
PMA / PMN Number
K082660
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

G4:10MAR2021. B4:13MAR2021. THE CUSTOMER EVALUATED THE DEVICE WITH ASSISTANCE FROM A PHILIPS REMOTE SERVICE ENGINEER (RSE), AND IT WAS DETERMINED THAT THE POWER SWITCH OVERLAY NEEDED TO BE REPLACED TO RETURN THE DEVICE TO WORKING CONDITION. THE CUSTOMER¿S BIO-MED REPLACED THE POWER SWITCH OVERLAY TO RESOLVE THE ISSUE AND BRING THE DEVICE BACK TO FUNCTIONALITY. A SIMILAR INVESTIGATION WAS PERFORMED WAS DETERMINED TO BE DELAMINATION AND CRACKING OF THE ENCAPSULATE EPOXY ENCASING THE LIGHT EMITTING DIODE (LED) DICE, AND PHYSICAL SHEARING OF THE EPOXY OR TRACE CAUSING OPENED AND INTERMITTED TRACE CONTINUITY TO THE LED. SUBMISSION OF A REPORT DOES NOT CONSTITUTE AN ADMISSION THAT MEDICAL PERSONNEL, USER FACILITY, IMPORTER, DISTRIBUTOR, MANUFACTURER, OR PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.

Additional Manufacturer Narrative · 1

DATE OF EVENT: (B)(6) 2020. DATE OF REPORT: 08JAN2021.

Description of Event or Problem · 1

IT WAS REPORTED TO PHILIPS THAT THE DEVICE HAD A LED (LIGHT EMITTING DIODE) FAILURE. THE DEVICE WAS NOT BEING USED ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO PATIENT OR USER HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
33979 V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE MNT V60

Patients

Seq Age Sex Outcome Treatment
1