Narratives

Additional Manufacturer Narrative · 1

NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT OCCLUSION ALARMS OCCURRED. SYSTEM CHECK WAS PERFORMED WITH TANDEM TECHNICAL SUPPORT AND IT WAS IDENTIFIED OCCLUSION WAS CAUSED BY A BLOCKAGE IN THE CARTRIDGE. CUSTOMER CHANGED PUMP SUPPLIES AND SUCCESSFULLY RESUMED INSULIN DELIVERY. CUSTOMER'S BLOOD GLUCOSE WAS 220-270 MG/DL AT TIME OF EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
35963 T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY ALTERNATE CONTROLLER ENABLED INFUSION PUMP QFG TANDEM DIABETES CARE 1000354 00850006613793

Patients

Seq Age Sex Outcome Treatment
1 63 YR INFUSION SET: TRUSTEELINSULIN: HUMALOG