8100 ALARIS PUMP MODULE
Report
- Report Number
- 2016493-2021-03187
- Event Type
- Malfunction
- Date Received
- January 8, 2021
- Report Date
- January 31, 2020
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- PMA / PMN Number
- K133532
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THE CUSTOMER REPORTED PROBLEM WAS CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS NOT RETURNED FOR SERVICING WHICH NOT CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT
BD QUALITY ADVOCATE, THIS NOTIFICATION IS TO INFORM YOU THAT A NEW CASE HAS BEEN CREATED WITH THE COMPLAINT TYPE CATEGORY INFUSION CA. CASE #: 00979462 CASE SUBJECT: NPI ERROR CODE 232.6040 ON 8100 UNIT ACCOUNT NAME: OHIOHEALTH MANSFIELD HOSPITAL ACCOUNT #: 1021989 ASSET NAME: 8015LS PCU DOM V9.33.1.2 A/B/G/N ASSET LOCATION: CONTACT: BEN JOHNSON CONTACT EMAIL: [email protected] CONTACT PHONE: CONTACT MOBILE: PATIENT OR USER INVOLVEMENT: NO PATIENT OR USER HARM: NO CASE DESCRIPTION: TECH CALLED IN FOR HELP ON 8100 UNIT SERIAL # (B)(4) FAILURE DEVICE TYPE: FAILURE PROBLEM TYPE: FAILURE MODE: CASE RESOLUTION: TECH CALLED IN FOR HLEP ON 8100 UNIT HAS ERROR CODE 232.6040, BEFORE CALL IN TECH SAID HE RAN PM TEST AND A SMALL INFUSE ERROR NEVER CAME BACK UP. EXPLAIN TO TECH THE ERROR HAS TO DO WITH THE DISPLAY FAILURE RATE DISPLAY SAFETY SYSTEM. IF THE ERROR COMES BACK UP THEY WILL NEED TO REPLACE THE DISPLAY MODULE ON THE UNIT. LOCATE BEHIND THE GREY PANEL OF THE FRONT DOOR. TECH THANK ME FOR MY ASSIST. REF:_00D30Y0YR._5000L1QKOKN:REF
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 38791 | 8100 ALARIS PUMP MODULE | PUMP, INFUSION | FRN | CAREFUSION SD | 8100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |