BINAX NOW COVID-19 AG CARD
Report
- Report Number
- 1221359-2021-00047
- Event Type
- Malfunction
- Date Received
- January 7, 2021
- Date of Event
- December 15, 2020
- Report Date
- March 25, 2021
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QKP
- PMA / PMN Number
- EUA202537
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ADDITIONAL INFORMATION: D4, H4, H6: INVESTIGATION TYPE/FINDINGS/CONCLUSION H10: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 130654 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-000 / LOT 130654 AND DEVICE PART NUMBER 195-430H / LOT 128930A. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 130654 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO THE SPECIFIC PATIENT SAMPLES.
REFERENCE MANUFACTURER REPORTS: 1221359-2021-00033, 1221359-2021-00034, 1221359-2021-00035, 1221359-2021-00036, 1221359-2021-00048, 1221359-2021-00056. THE INVESTIGATION IS STILL IN PROGRESS, A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.
THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE BINAX NOW COVID-19 AG CARD OVER A PERIOD OF 3 WEEKS INVOLVING TWO PATIENTS. THIS REPORT ADDRESSES PATIENT ONE(1) OF TWO (2) RELATED TO THE EVENT RELATED LOT NUMBER 130654 (ADDITIONAL TESTING). THE CUSTOMER REPORTED A FALSE POSITIVE RESULT DURING ADDITIONAL TESTING OF PATIENT 1 ON (B)(6) 2020. THE CUSTOMER DID NOT PROVIDE SAMPLING DETAILS PERTAINING TO REPEAT OR CONFIRMATION TESTING. THE CUSTOMER STATED THE PATIENT WAS NOT SYMPTOMATIC. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT. PER BINAXNOW COVID-19 AG CARD PRODUCT INSERT INTENDED USE: POSITIVE RESULTS INDICATE THE PRESENCE OF VIRAL ANTIGENS, BUT CLINICAL CORRELATION WITH PATIENT HISTORY AND OTHER DIAGNOSTIC INFORMATION IS NECESSARY TO DETERMINE INFECTION STATUS. POSITIVE RESULTS DO NOT RULE OUT BACTERIAL INFECTION OR CO-INFECTION WITH OTHER VIRUSES. THE AGENT DETECTED MAY NOT BE THE DEFINITE CAUSE OF DISEASE. PER BINAXNOW COVID-19 AG CARD PRODUCT INSERT INTENDED USE (LIMITATIONS): FALSE RESULTS MAY OCCUR IF SPECIMENS ARE TESTED PAST 1 HOUR OF COLLECTION. SPECIMENS SHOULD BE TESTED AS QUICKLY AS POSSIBLE AFTER SPECIMEN COLLECTION. DUE TO THE RISK OF A FALSE POSITIVE RESULT LEADING TO POSSIBLE EXPOSURE TO COVID-19 THROUGH THE QUARANTINE OF THE FALSE POSITIVE PATIENT WITH TRUE POSITIVE PATIENT(S), DELAYED EMERGENCY TREATMENT AND/OR INAPPROPRIATE TREATMENT WITH ANTIVIRAL THERAPY, THE INCIDENT SHALL BE CONSIDERED REPORTABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 32058 | BINAX NOW COVID-19 AG CARD | LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 | QKP | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 130654 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |