DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2021-003284
- Event Type
- Malfunction
- Date Received
- January 7, 2021
- Date of Event
- December 15, 2020
- Report Date
- January 11, 2021
- Manufacturer
- DEXCOM, INC.
- Product Code
- QBJ
- UDI-DI
- 00386270000385
- PMA / PMN Number
- DEN170088
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
COM-(B)(4). 3004753838-2021-003284 WAS REPORTED IN ERROR. PLEASE DISREGARD INITIAL REPORTING OF THIS EVENT AS THIS EVENT HAS NOW BEEN DEEMED NOT REPORTABLE.
SUBSEQUENT TO THE INITIAL MDR, IT WAS DETERMINED THAT A REPORT WAS SUBMITTED IN ERROR. UPON FURTHER REVIEW, IT WAS DETERMINED THAT THE PATIENT ALLEGATION DOES NOT MEET THE CRITERIA OF A REPORTABLE EVENT.
(B)(4).
IT WAS REPORTED THAT A PAIRING FAILURE OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE A FAILED TRANSMITTER ERROR. IN ADDITION, TRANSMITTER FAILED ERROR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE REPORTED EVENT OF A PAIRING FAILURE IS REPORTABLE BASED ON THE FINDING OF A TRANSMITTER FAILED ERROR. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 26662 | DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | QBJ | DEXCOM, INC. | 9445-02 | 7275538 | 00386270000385 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |