FDA Adverse Event Malfunction Summary report: N

VSM

MDR report key: 1113326 · Received August 12, 2008

Report

Report Number
3023750-2008-00192
Event Type
Malfunction
Date Received
August 12, 2008
Date of Event
July 15, 2008
Report Date
July 15, 2008
Manufacturer
WELCH ALLYN PROTOCOL, INC.
Product Code
DQA
PMA / PMN Number
K031740
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION - INSUFFICIENT EVIDENCE OR INFORMATION LEADING TO CAUSE CANNOT BE DETERMINED. EVALUATION SUMMARY: THE DEVICE IS AN AC TO DC POWER CONVERTER THAT HAS NOT BEEN RETURNED BUT WAS INVESTIGATED USING PHOTOGRAPHIC IMAGES. THE IMAGES REVEALED THAT THE OUTER ORANGE INSULATION ON THE AC POWER CORD (RESULT) WAS STRETCHED AND SEPARATED, REVEALING THE INTERNALLY INSULATED WIRES WHICH WERE UNAFFECTED BY THE STRETCHING EVENT. EVALUATING THE PICTURES TAKEN OF THE AFFECTED MATERIAL CONFIRMED THE STRETCHED AND SEPARATED OUTER INSULATION SHEATH OF THE AC POWER CORD. THE CAUSE OF THE STRETCHING ACTION TO ALLOW THE OUTER INSULATION TO BREAK CANNOT BE DETERMINED.

Description of Event or Problem · 1

THE CUSTOMER RAISED SAFETY CONCERNS RELATING TO THE OUTER ORANGE INSULATION ON THE 240V AC POWER CORD. THE OUTER ORANGE INSULATION STRETCHED AND SEPARATED SEVERAL INCHES BELOW THE IEC CONNECTOR ALLOWING THE INDIVIDUAL INSULATED AC CONDUCTOR WIRES TO BE VIEWED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VSM DQA WELCH ALLYN PROTOCOL, INC. VSM300

Patients

Seq Age Sex Outcome Treatment
1