FDA Adverse Event Malfunction Summary report: N

SIMPLEX P-JAPANESE TWIN PACK

MDR report key: 1113309 · Received August 12, 2008

Report

Report Number
9610726-2008-00050
Event Type
Malfunction
Date Received
August 12, 2008
Date of Event
July 17, 2008
Report Date
July 17, 2008
Manufacturer
STRYKER ORTHOPAEDICS LIMERICK
Product Code
LOD
PMA / PMN Number
N17004
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN EVAL OF THE DEVICE CANNOT BE PERFORMED AS THE DEVICE WAS NOT RETURNED TO THE MFR. IF ADD'L INFO BECOMES AVAILABLE, IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SURGERY THE CEMENT SET TOO QUICKLY. IT SET IN 6 TO 7 MINUTES ALTHOUGH USUALLY 11 TO 12 MINUTES. THE SURGEON IS VERY EXPERIENCED WITH THE SIMPLEX CEMENT AND HE ADDRESSED THAT HE KEPT AND HANDLED THE CEMENT AS USUAL HOWEVER, IT SET TOO QUICKLY. OPERATING ROOM TEMPERATURE WAS AT 24C. DISTRIBUTORS' WAREHOUSE TEMPERATURE WAS 25C DURING THE DAY. HOWEVER, THE AIR CONDITIONER WAS OFF DURING NIGHTS. THE TEMPERATURE AT NIGHT IS UNK EXACTLY; HOWEVER, IT GETS VERY WARM DURING SUMMER TIME (AROUND 30 C OR HIGHER). THE MIXING WAS DONE MANUALLY. NO ANTIBIOTIC WAS MIXED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIMPLEX P-JAPANESE TWIN PACK IMPLANT LOD STRYKER ORTHOPAEDICS LIMERICK NA JEP019

Patients

Seq Age Sex Outcome Treatment
1 UNK Other