FDA Adverse Event
Malfunction
Summary report: N
DIMENSION CLINICAL CHEMISTRY SYSTEM
MDR report key: 1113284
·
Received August 8, 2008
Report
- Report Number
- 1226181-2008-00095
- Event Type
- Malfunction
- Date Received
- August 8, 2008
- Date of Event
- July 19, 2008
- Report Date
- July 19, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JJE
- PMA / PMN Number
- K010061
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A SIEMENS HEALTHCARE DIAGNOSTICS INC FIELD SERVICE REPRESENTATIVE (FSE) WAS DISPATCHED TO THE ACCOUNT ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS A SLIPPING HM PUMP. THE FSE CORRECTED THE MALFUNCTION. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
Description of Event or Problem · 1
A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED AND A NEGATIVE RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | JJE | SIEMENS HEALTHCARE DIAGNOSTICS INC. | XPANDHM | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |