FDA Adverse Event
Malfunction
Summary report: N
LIFECARE PLUM
MDR report key: 1113083
·
Received August 1, 2008
Report
- Report Number
- 1113083
- Event Type
- Malfunction
- Date Received
- August 1, 2008
- Date of Event
- June 14, 2008
- Report Date
- July 28, 2008
- Manufacturer
- ABBOTT LABORATORIES, INC.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- TN, US
Narratives
Description of Event or Problem · 1
PLUM PUMP HAD A MALFUNCTION CODE OF E24. RN TOOK PUMP OUT OF SERVICE AND TAGGED IT. AN ERROR E24 INDICATES A PROBLEM WITH THE BATTERY. THE BATTERY WAS REPLACED AND THE PUMP RETESTED. THE PUMP CHECKED OUT TO BE OKAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFECARE PLUM | PUMP, IV | FRN | ABBOTT LABORATORIES, INC. | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |