FDA Adverse Event Injury Summary report: N

STEM: QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 3

MDR report key: 11129987 · Received January 7, 2021

Report

Report Number
3005180920-2021-00019
Event Type
Injury
Date Received
January 7, 2021
Date of Event
December 10, 2020
Report Date
January 7, 2021
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030802157
PMA / PMN Number
K082792
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON (B)(6) 2020: LOT 161558: 40 ITEMS MANUFACTURED AND RELEASED ON 20-MAY-2016. EXPIRATION DATE: 2021-05-09. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 37 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. PRELIMINARY INVESTIGATION PERFORMED ON (B)(6) 2020 BY MEDACTA R&D HIP PROJECT MANAGER: FROM THE RECEIVED IMAGES IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF THE EVENT. IT MAY BE POSSIBLE THE PRESENCE OF RADIOLUCENCIES AROUND THE STEM IN THE X-RAY IMAGE, WHILE THE EXPLANTED STEM DID NOT SHOW ANY PARTICULAR SIGN. A MORE DETAILED ANALYSIS CAN BE DONE WITH THE PHYSICAL PIECE. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER: FEMORAL COMPONENT REVISION PERFORMED 4 YEARS AFTER PRIMARY CEMENTLESS TOTAL HIP ARTHROPLASTY IN (B)(6) YEAR OLD MAN. NO INFORMATION CONCERNING PATIENT GENERAL HEALTH STATUS AND THE PRESENCE OF COMORBIDITIES IS AVAILABLE. IN THE RADIOGRAPHIC IMAGE PROVIDED, RADIOLUCENT LINES AND SIGNS OF STRESS SHIELDING ARE VISIBLE. ASEPTIC LOOSENING IS A POSSIBLE LITERATURE DESCRIBED ADVERSE EVENT AFTER PRIMARY CEMENTLESS HIP ARTHROPLASTIES AND CAUSES ARE OFTEN UNKNOWN. THE REASON OF THIS FAILURE CANNOT BE DETERMINED.

Description of Event or Problem · 1

REVISION SURGERY 4 YEARS AND 3 MONTHS AFTER PRIMARY FOR STEM LOOSENING. QUADRA H REPLACED WITH QUADRA C. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27859 STEM: QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 3 CEMENTLESS HIP STEM JDI MEDACTA INTERNATIONAL SA 01.12.033 161558 07630030802157

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention